background image

 

Operation 

TDS measuring device 

 
The battery powered meter shows the TDS (total dissolved 
solids) in ppm in the water. See Fig. 2. 

Operation:

 Switch on using the red button on the top left, 

then select the measuring point, (inlet or outlet). As soon as 
the value on the output is > 20 ppm the resin in the cartridge 
is exhausted. The untreated water hardness can be 
determined by measuring the value at the inlet and the 
capacity of the resin can be calculated off the capacity graphs 
(Fig. 3). 
 
In order to conserve batteries, the meter turns off 
automatically after 20 seconds. Reset if necessary. 
 

Flow and quantity meter 

 
The volume of demineralized, and total water can be read off 
the flow meters. 

Operation:

 The places after the comma show the litres, the 

places in front of the comma show the cubic metre (1,000 
litres). Example: 0001.248 is 1,248 L (3.8 L = 1 US gal.). 
The top flow meter shows the quantity of completely 
demineralized water, the bottom meter shows the total 
amount of water, including mixed water. 
 
It is recommended to log the last meter reading when 
changing the resin because the meter cannot be reset. 
 

Mixing valve 

 
If demineralization only suffices up to a certain hardness, 
then untreated water can be added. The mixing valve is used 
as well when cleaning the closed loop hydronic system: 
Rinse with untreated water and then fill the closed loop 
hydronic system with demineralized water by moving the 
valve lever. 
 

Pressure reducing valve 
 

The pressure reducing valve regulates the pressure of the 
untreated water with regard to the system.  

Operation:

 Use the screw driver to turn the adjusting screw 

in the center of the valve: To the left: pressure is reduced; to 
the right, pressure is increased. The higher the pressure, the 
faster the water flows through the fill-up station. The 
adjusted pressure can be read off the pressure gauge. 

 
The pressure may not be adjusted higher than what the pressure the relief valve of the closed loop 
hydronic system is set to. In case of doubt, do not set higher than 30 psi (207 kPa). 

Capacity 6.6 gpm 
(25 L/min) 

Fig. 2: Operation 

(29-87 psi) 

Содержание Puropal-Pro-25

Страница 1: ...or drinking water The filling apparatus is equipped with precise meters for monitoring pure water production in terms of quality and quantity If the mixed bed ion exchange resin capacity is exhausted...

Страница 2: ...ection under pressure is not permitted Under normal domestic water pressure the flow rate through the Puropal Jr Pro is approximately 6 6 gpm 25 l min Figure 1 Connections Open the drain valve on the...

Страница 3: ...meter shows the total amount of water including mixed water It is recommended to log the last meter reading when changing the resin because the meter cannot be reset Mixing valve If demineralization...

Страница 4: ...below Example With a hardness of 11 7 gpg 200 ppm Puropal Pro 25 provides 725 gal 2750 L of completely demineralized water and Puropal Pro 50 provides 1780 gal 6750 L Capacity of the Puropal Pro 25 5...

Страница 5: ...connected to the untreated water tap 3 2 Connect hose to drain valve 4 and insert it into the supplied water permeable collection sack Tightly connect the sack to the top of the hose with a cable bin...

Страница 6: ...l leak into the system Figure 5 Filling the Puropal Pro with new resin Check that the spray hose is pressed tightly into the head and is sealed otherwise resin will leak into the system Make sure that...

Страница 7: ...to read or contains illegible characters Change batteries See Fig 6 below The display shows ERR Measuring device malfunction Reset by switching on and off quickly 2 3 times until the display is norma...

Страница 8: ...de of the cartridge is blocked Loosen screw rinse out fine sieve and reattach See Fig 7 below Figure 7 Location of fine sieve at outlet Drain sieve for residual water Damage Preventative solution The...

Отзывы: