2.2.4.
AC Current (ACA)
Range
Accuracy
Resolution
200 mA
± (2,0 % + 5)
100 µA
2 A
± (3,0 % + 5)
1 mA
20 A
± (3,0 % + 10)
10 mA
Max. measuring volt drop: 200mV
Max. input current: 20A ( the test time should be in 10 seconds)
Overload protection: 2A/250 V quick-action fuse, 20A un-fused
Frequency response: (40
~
200) Hz
Display: sine wave RMS (mean value response)
2.2.5.
Resistance (
Ω
)
Range
Accuracy
Resolution
200 Ω
± (0,8 % + 5)
0,1 Ω
2 kΩ
± (0,8 % + 3)
1 Ω
20 kΩ
10 Ω
200 kΩ
100 Ω
2 MΩ
1 kΩ
200 MΩ
± [5,0 %(RDG -10) +20]
100 kΩ
Open Voltage: less than 3V
Overload protection: 250V DC and AC peak value
NOTE:
A.
At 200
Ω range, should make the test leads short, and measure the wire resistance, then subtract
from the actual measuring.
B.
At 200M Ω range, should make the test leads short, the meter will display 1.0M Ω. It’s normal and
has no effect on the accuracy. Should be subtracted from actual measuring.