Range 

Accuracy 

Resolutin 

400

µ

±(1.0% + 10 d) 

0.1

µ

4000

µ

1

µ

40mA 

10

µ

400mA 

100

µ

10A 

±(1.2% + 10 d) 

10mA 

 

Max measurement voltage drop: Full Range mA is 0.4V, A is 100mV; 

Max input current : 10A (less than 15 seconds); 

Overload Protection: 0.4A/250V restorable fuse, 10A/250V fuse. 

2.2.4.

 

ACA 

1.

 

Insert the black meter pen into the “COM” input terminal and red one into the 

 input 

terminal. (Max 400mA), or 10A input terminal (Max 10A) 

2.

 

Rotate function switch to Current gear. Press “SELECT” button to select the AC measurement 

mode. Then connect the test leads to the tested circuit in serial, the tested current value and the 
current polarity of the point where the red one is connected will be displayed on the screen 
simultaneously. 

Caution: 

1.

 

If “OL” is displayed on LCD, it indicates the tested current value has exceeded the present range 

limit, please select higher range to complete the measurement. 

2.

 

The Max input value is 400mA or 10A (Depending on the terminal where the red test lead is 

contacted). The overrated current will lead to fuse melt or even damage the meter. 

Range 

Accuracy 

Resolution 

400uA 

±(1.5% + 10 cyfr) 

0.1uA 

4000uA 

1uA 

40mA 

10uA 

400mA 

100uA 

10A 

±(2.5% + 15 cyfr) 

10mA 

Max measurement voltage drop: Full Range mA is 0.4V, A is 100mV; 

Max input current: 10A (less than 15 seconds) 

Содержание AX-105

Страница 1: ...DIGITAL MULTIMETER AX 105 OPERATION MANUAL...

Страница 2: ...r immediately Digital Multimeter 1PC Instruction Manual 1 copy Test Leads 1 pair Temperature Probe K Thermocouple 1 PC AAA 1 5v 7 battereies 2 PCs Holster 1 PC III Safety Considerations The design of...

Страница 3: ...ultimate voltage value permitted for measurement is marked on the instrument panel and never measure the value exceeding the standard Don t input the ultimate value out of regulation in order to avoi...

Страница 4: ...uty cycle could be selected by pressing this button in AC DC voltage or AC DC current gear 3 2 SELECT Function Switch Press this button the function could switch between DC AC 3 3 REL Relative Value M...

Страница 5: ...ssing the HOLD button to turn on the instrument VI Property 1 General Property 1 1 Display LCD 1 2 Max Display 3999 3 counts automatic polarity display and unit display 1 3 Measuring Method Dual integ...

Страница 6: ...aution 1 Don t measure voltage over 600V Otherwise ther is a danger of meter being damaged 2 When measuring high voltage special attention should be given to personal safety and avoid your body gettin...

Страница 7: ...age value response RMS of sine wave 2 2 3 DCA 1 Insert the black test lead into the COM input terminal and red one into the input terminal Max 400mA or 10A input terminal Max 10A 2 Rotate function swi...

Страница 8: ...leads to the tested circuit in serial the tested current value and the current polarity of the point where the red one is connected will be displayed on the screen simultaneously Caution 1 If OL is di...

Страница 9: ...ent range limit please select higher range to complete the measurement When measuring the Resistor higher than 1M the instrument will take several seconds to make the reading stable It is normal when...

Страница 10: ...the result does not meet the above it means the diode is bad 6 Press SELECT button to select the Continuity measurement mode 7 Connect the meter pens to two points of the tested circuit If the built i...

Страница 11: ...e 2 2 8 Frequency F 1 Connect test leads and shielded cable to COM terminals 2 Rotate function switch to Hz gear Connect test leads and the cable to the signal source or the tested load The tested sig...

Страница 12: ...put the working end temperature measurement end of thermocouple on the surface or inside the object to be tested Then you can read temperature from the screen and the data is in Centigrade Caution 1...

Страница 13: ...batteries and replace them with two new ones Although a 1 5V battery of any standard can be used but in order to lengthen the operation life alkaline batteries should be used 3 Mount the battery door...

Страница 14: ...ct and in case readers find any errors and missing parts please contact the manufacturer The Company shall not be held liable for any accidents and hazards resulted from the mal operations by the user...

Отзывы: