7
4. Make the test leads connected to the testing point. The voltage of the point where the red pen is
contacted will be displayed on the screen.
Caution:
1. Don’t measure voltage over 600V. Otherwise, there is a danger of meter being damaged.
2. When measuring high voltage, special attention should be given to personal safety and avoid your
body getting in touch with high voltage circuit.
Range
Accuracy
Resolution
4V
±(0.8% + 6 d)
1mV
40V
10mV
400V
100mV
600V
±(1.0% + 6 d)
1V
Input Impedance: >10M
Ω
Overload protection: 600V DC or 600V AC Peak Value
Frequency Response: (50
~
200) Hz
Display: Average value response (RMS of sine wave)
2.2.3.
DCA
1.
Insert the black test lead into the “COM” input terminal and red one into the
input
terminal. (Max 400mA), or 10A input terminal (Max 10A).
2.
Rotate function switch to Current gear. The initiate state of the meter is in automatic range
status, which shows “DC” symbol. Then connect the test leads to the tested circuit in serial,
the tested current value and the current polarity of the point where the red one is contacted will
be displayed on the screen simultaneously.
Caution:
1.
If “Ol” is displayed on LCD, it indicates the tested current value has exceeded the present range
limit, please select higher range to complete the measurement.
2.
The Max input value is 400mA or 10A. (Depending on the terminal where the red test lead is
contacted).