4) Read the temperature in the LCD display.
5) Push SEL button to shift the unit of ℃ or °F

6.6. NCV (Non-Contact Voltage) Detection

WARNING:

Due to external interference source, this function may cause wrong voltage detection, the detection result
is for reference only.

Set the rotary switch to NCV position, contact the top part of meter with the circuit under test,
the indicating LED will flash and audible signal will sound.

NOTE:
� The detection result is for reference, do not determine the voltage by NCV detection ONLY.
� Detection may interfere by socket design, insulation thickness and other variable conditions.
� The external interference sources, such as flashlight, motor, etc, may cause the wrong detection.

6.7. LINE (Live Wire Recognition) Test

Set the rotary switch to LINE position, connect the black test lead to COM jack and red test lead to VΩmA
jack, hold the insulation part of black test lead and not put into circuit under measurement; contact the
red test lead to live wire, the buzzer of meter will be activated and red LED will be flickered, when the red
test lead connect the earth line, the buzzer does not sound and LED will not flicker.
NOTE:
When the circuit is in serious leakage (approx. over 15V), the red test lead even contact earth line, the
buzzer of meter will be sounded and LED will be flickered.

6.8. Display Backlit

Press the

button for 1 or 2 seconds to turn on or off the display backlit function, the backlight will

automatically turn off after 10 seconds.

7. MAINTENANCE

7.1.

WARNING:

To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover
or the battery or fuse covers.

Содержание AX-102

Страница 1: ...AX 102 User s Guide 1...

Страница 2: ...perature Battery and Non Contact Voltage Detection Proper use and care of this meter will provide many years of reliable service To fully utilize this meter please keep this manual for reference caref...

Страница 3: ...and the meter itself for any damage before operating the meter Do not measure voltage if the voltage on the terminals exceeds 1000V above earth ground Use great care when making measurements if the v...

Страница 4: ...4 PANEL DESCRIPTION 1 LCD display 2 Select Button shift the measurements of AC DC current Temperature Diode Continuity check 3 Range button Auto Manual range shift 4 Rotary switch...

Страница 5: ...Operating Temperature 0 40 32 F 104 F Operating Humidity 80 RH Storage Temperature 10 60 14 F 122 F Storage Humidity 70 RH Power Supply 1 5V Battery x 2pcs Dimension 144 x 70 x 32mm Weight Approx 200...

Страница 6: ...ange 40 400Hz 5 2 4 DC Current Range Resolution Accuracy 200 A 0 1 A 1 0 reading 5 digits 2mA 1 A 1 0 reading 5 digits 20mA 10 A 1 0 reading 5 digits 200mA 100 A 2 0 reading 5 digits 2A 1mA 3 0 readin...

Страница 7: ...50V DC or 250V AC RMS 6 OPERATING INSTRUCTIONS 6 1 AC and DC Voltage Measurement WARNING Risk of electrocution High voltage circuits both AC and DC are very dangerous and should be measured with great...

Страница 8: ...into the mA jack for current measurement between 200mA to 10A insert the red test lead banana plug into 10A jack 3 Break the circuit under test connect the black test lead to the more negative side of...

Страница 9: ...est of the circuit will not interfere with the resistance reading 4 Read the resistance in the LCD display NOTE The measured value of a resistor in a circuit usually is different from the rated value...

Страница 10: ...re Recognition Test Set the rotary switch to LINE position connect the black test lead to COM jack and red test lead to V mA jack hold the insulation part of black test lead and not put into circuit u...

Страница 11: ...er 2 Open the rear battery cover by using screwdriver 3 Insert the battery into battery holder observing the correct polarity 4 Put the battery cover back in place secure with the screws 7 3 Replacing...

Отзывы: