I. Operation

Place the scale on a flat and stable surface.

1. Press ON/OFF key.

2. Wait until [0.0] is displayed.

3. Put the object(s) on the weighing platform.

4. Using the MODE key, you can switch between the weighing units.

II. Tare Weighing.

1. Turn on scale as described above.

2. Place the “tare item” on the platform.

3. Press TARE and wait until [0.0] is displayed.

4. Add the “net-weight-item”.

III. Add-Up Feature

1. In normal weighing mode, place an object on the platform 

2. Once weight is displayed, press “ADD UP” key and replace item with 

another and press “ADD UP” again

3. The scale will display the sum of the two weights.

4. You may repeat until scale reaches max capacity ( display will show 

“0”  when capacity is exceeded )

IV. Calibration

1. Turn the scale on as described above and wait for the display to 

show “0.000g”. 

2. Press and hold the TARE key until the display shows “CAL” then a 

random number will be displayed (ISN number).

3. Make sure no weight is on the platform and press the TARE key 

again. The display will flash CAL followed by “0.000g”.

4. Now place a 10g calibration weight on the platform and press TARE 

again when the number stabilizes. The display will flash “10.000g” 

followed by the ISN number again.

5. At this point, place another 10g on the platform to make a total of 

20g on the scale and press TARE when the number stabilizes.

6. The display will show “20.000g” followed by “PASS”. Press ON/OFF 

Отзывы: