1010 Professional Transducer Display
Quick Start
Page 9
Advanced Witness Systems Ltd
1010 Professional Transducer Display Operating Manual
If “
Yes
” is selected the PTD will use the settings selected at its last
power down. If “
No
” is selected the user will then be prompted to
select 1
st
peak auto cancel, and/or set limits, units of
measurement, and the mode of operation.
1
st
PEAK AUTO CANCEL PROMPT SET LIMIT PROMPT
UNITS
The PTD uses 5 Units of measurement
N.m, cN.m, Lbf.Ft, Lbf.In
and Ozf.In
. These are selected for the displayed torque value and
setting the limits. Selected units can be changed from the main
screen by pressing back twice and reselecting a new unit.
UNIT SELECTION. UPPER LIMIT SET
LIMITS
Set the “
target torque limit
” to the required setting, see key chart
on page 12 on how to use buttons for limit value setting.
Set the “
upper limit
” to the maximum permissible torque reading.
Please note this is limited to 110% of the transducers full scale torque
value and must be above the target torque or the “Over Range” warning
will not allow you to proceed to the next screen
.
Set the
“
Lower limit
”
to the minimum permissible torque reading.
Please note this must be set below the target torque or the “Over
Range” warning will not allow you to proceed to the next screen
.