1010 Professional Transducer Display
Calibration Guide
Page 13
Advanced Witness Systems Ltd
1010 Professional Transducer Display Operating Manual
CALIBRATION GUIDE
The AWS PTD unit can be used in conjunction with a standard to
calibrate a torque transducer.
BUTTONS Functions (as shown on screen)
INC
Increases the selected number, cycles through zero.
Changes selected units.
>/TARE
Pressing moves the selected digit to the right, this
cycles back through. Pressing and holding the button
will tare/zero the reading.
CAL
This will calibrate the transducer to the selected torque
range.
STORE
Will store the selected torque range.
ORDER OF CALIBRATION
1. Fit transducer in appropriate Calibration machine.
2. Attach the transducer to the PTD via cable.
3. Place the PTD into Calibration Mode by Pressing and
holding the furthest left and furthest right soft key buttons
whilst turning the PTD on. This will activate the Calibration
Mode.
4. Allow the display and transducer to thermally stabilise for 5
minutes.
5. Use the “
INC
” and “
>/TARE
” buttons to change the N.M.
RANGE to the maximum range of the transducer. Press
“
STORE
” to store this number.
6. Use the “
INC
” and “
>/TARE
” buttons to change the units to
the required units. Press “
STORE
” to store this number.
7. Press and hold “
>/TARE
”.