When triggering externally, the instrument can be set such that the output pulse
width tracks the pulse width on this input, or the output pulse width can be set
independently.
6. GPIB Connector
. A standard GPIB cable can be attached to this connector to allow
the instrument to be computer-controlled. See the “Programming Manual for -B
Instruments” for more details on GPIB control.
7. RS-232 Connector.
A standard serial cable with a 25-pin male connector can be
attached to this connector to allow the instrument to be computer-controlled. A user
name (“admin”) and a password (“default”, as shipped from the factory) are required
when logging into a serial terminal session. The internal controller attempts to auto-
sense the parity setting. It may be necessary to send a few return characters before
attempting a login in order to provide enough data to allow this auto-sensing to work.
(A standard Linux “agetty” process is used to implement serial control internally.) See
the “Programming Manual for -B Instruments” for more details on RS-232 control.
8. Network Connector
. (Optional feature. Active on -VXI units only.) This Ethernet
connector allows the instrument to be remotely controlled using the VXI-11.3, ssh
(secure shell), telnet, and http (web) protocols. See the “Programming Manual for -B
Instruments” for more details.
19
Содержание AVR-G5-B
Страница 24: ...Use the correct load impedance 100 k and 100 pF Keep output cable lengths to 2 meters 6 feet or less 24...
Страница 34: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Страница 35: ...PCB 235B HIGH VOLTAGE POWER SUPPLY...
Страница 37: ...PCB 228B OP AMP...
Страница 38: ...PCB 287A TTL BUFFER...
Страница 43: ...PERFORMANCE CHECK SHEET 43...