GENERAL INFORMATION
BASIC PULSE CONTROL
This instrument can be triggered by its own internal clock or by an external TTL trigger
signal. In either case, two output channels respond to the trigger: OUT and SYNC.
OUT. This is the main output.
SYNC. The SYNC pulse is a fixed-width TTL-level reference pulse used to
trigger oscilloscopes or other measurement systems. When the delay is set to a
positive value the SYNC pulse precedes the OUT pulse. When the delay is set to
a negative value the SYNC pulse follows the OUT pulse.
These pulses are illustrated below, assuming internal triggering and a positive delay:
Figure A
If the delay is negative, the order of the SYNC and OUT pulses is reversed:
Figure B
19
SYNC OUT
(generated by the
internal oscillator)
100 ns, FIXED
MAIN OUTPUT
PULSE WIDTH
DELAY > 0
AMPLITUDE,
VARIABLE
3V, FIXED
SYNC OUT
(generated by the
internal oscillator)
100 ns, FIXED
MAIN OUTPUT
PULSE WIDTH
DELAY < 0
AMPLITUDE,
VARIABLE
3V, FIXED
Содержание AVP-AV-HV3-B
Страница 32: ...WIRING DIAGRAMS WIRING OF AC POWER ...
Страница 33: ...WIRING OF HIGH VOLTAGE POWER SUPPLY ...
Страница 34: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY ...
Страница 35: ...PCB 235C HIGH VOLTAGE DC POWER SUPPLY ...
Страница 36: ...PCB 231A INTERNAL OFFSET GENERATOR ...
Страница 37: ...PCB 216F PW PRF TWEAK ...
Страница 38: ...PCB 151D RANGE POLARITY RELAY BOARD ...
Страница 39: ...PCB 241B DELAY ...
Страница 40: ...PCB 313A AUXILIARY DACS ...
Страница 41: ...PCB 335A TWO CHANNEL ENCODER ...
Страница 42: ...PCB 104G KEYPAD DISPLAY BOARD ...
Страница 46: ...MAIN WIRING 2CHX UNITS ...