4000
4-/5-Gas Measuring Instruments Unit
Service Manual
5-13
N.B.
Large amounts of deposits tend to be washed away when you first start
cleaning. Do not therefore feed the cleaning fluid back into the circuit to
start with. The length of time depends on the level of soiling; guideline
~30 s).
•
Switch the instrument on and activate the pump.
•
The cleaning process itself should take about half an hour
•
When cleaning is completed, rinse the system in the same way with cold water.
•
Remove the cleaning hose from the check valve and let the pump run on for ~30 s.
•
Switch the pump off.
•
Remove the cleaning hose and the probe hose.
•
Clean the probe hose and probe with clean, oil-free compressed air.
Caution: the probe hose must not be connected to the exhaust gas tester!
•
Replace the condensate filter on the rear panel of the instrument.
•
Connect the hose between the condensate filter and check valve.
•
Replace the filter in the probe hose.
•
After cleaning, carry out a leak check.
Содержание AVL DiCom 4000
Страница 4: ......
Страница 6: ......
Страница 22: ...Periodic Maintenance AVL 4000 1 12 Service Manual ...
Страница 47: ...4000 Evaluation Unit Service Manual 4 15 4 9 Main Board 4 9 1 Components Location Diagram ...
Страница 48: ...Evaluation Unit AVL 4000 4 16 Service Manual 4 9 2 Circuit Diagrams ...
Страница 49: ...4000 Evaluation Unit Service Manual 4 17 ...
Страница 50: ...Evaluation Unit AVL 4000 4 18 Service Manual ...
Страница 51: ...4000 Evaluation Unit Service Manual 4 19 ...
Страница 52: ...Evaluation Unit AVL 4000 4 20 Service Manual ...
Страница 53: ...4000 Evaluation Unit Service Manual 4 21 4 10 Speed Angle Board 4 10 1 Components Location Diagram ...
Страница 54: ...Evaluation Unit AVL 4000 4 22 Service Manual 4 10 2 Circuit Diagrams ...
Страница 55: ...4000 Evaluation Unit Service Manual 4 23 ...
Страница 56: ...Evaluation Unit AVL 4000 4 24 Service Manual ...
Страница 57: ...4000 Evaluation Unit Service Manual 4 25 ...
Страница 58: ...Evaluation Unit AVL 4000 4 26 Service Manual 4 11 Pneumatics Board 4 11 1 Components Location Diagram ...
Страница 59: ...4000 Evaluation Unit Service Manual 4 27 4 11 2 Circuit Diagram ...
Страница 60: ...Evaluation Unit AVL 4000 4 28 Service Manual 4 12 Printer Adapter 4 12 1 Components Location Diagram ...
Страница 61: ...4000 Evaluation Unit Service Manual 4 29 4 12 2 Circuit Diagram ...
Страница 62: ...Evaluation Unit AVL 4000 4 30 Service Manual 4 13 Memory Adapter 4 13 1 Components Location Diagram ...
Страница 63: ...4000 Evaluation Unit Service Manual 4 31 4 13 2 Circuit Diagram ...
Страница 64: ...Evaluation Unit AVL 4000 4 32 Service Manual 4 14 Connector Adapter 4 14 1 Components Location Diagram ...
Страница 65: ...4000 Evaluation Unit Service Manual 4 33 4 14 2 Circuit Diagram ...
Страница 67: ...4000 Evaluation Unit Service Manual 4 35 4 15 2 Circuit Diagram ...
Страница 69: ...4000 Evaluation Unit Service Manual 4 37 4 16 2 Circuit Diagram ...
Страница 71: ...4000 Evaluation Unit Service Manual 4 39 4 17 2 Circuit Diagram ...
Страница 73: ...4000 Evaluation Unit Service Manual 4 41 4 18 2 Circuit Diagram ...
Страница 75: ...4000 Evaluation Unit Service Manual 4 43 4 19 2 Circuit Diagram ...
Страница 77: ...4000 Evaluation Unit Service Manual 4 45 4 20 2 Circuit Diagram ...
Страница 79: ...4000 Evaluation Unit Service Manual 4 47 4 21 2 Circuit Diagram ...
Страница 124: ...4 5 Gas Measuring Instruments Unit AVL 4000 5 28 Service Manual ...
Страница 149: ...4000 Opacity Measuring Chamber Type A Service Manual 6 25 Fig 6 30 Fig 6 31 ...
Страница 150: ...Opacity Measuring Chamber Type A AVL 4000 6 26 Service Manual Fig 6 32 ...
Страница 159: ...4000 Opacity Measuring Chamber 4000 Service Manual 7 3 Mainboard Chamber Adapter BB0814776 DS0992 BV1982 Fig 7 3 ...
Страница 180: ...Spare Parts Lists AVL 4000 8 14 Service Manual ...
Страница 182: ...Tools for the Service Technician AVL 4000 9 2 Service Manual ...