Maintenance
AVL 415SE Smoke Meter — Maintenance Guide
Sealing ring
1.
Check the sealing rings for damage or deposits of paper fibres.
2.
Remove paper fibres with a soft, damp cloth.
3.
Replace sealing rings if damaged.
Never re-glue used sealing rings.
4.
Use only the defined adhesive for glueing the sealing rings.
Carefully follow the instructions of the glue manufacturer for the usage of
the glue.
5.
Close the clamping lever (remember the hose clamps).
Gasket 30mm 4152X38./C00.0
Material number: YM3499SP
Mounting ring
We recommend ordering this part for fixing the gasket (material number
YM3499SP) into the suction unit of the clamping and drive mechanism:
Mounting ring 30 mm 4152X41./C00.0
Material number: YM3516SP
NOTICE
Apply only a thin film of lubrication - only into region between the upper and
lower clamping piece (
), so that no oil can reach the sealing rings.
Never clean the elastomer parts (sealing rings) in an ultrasonic bath.
Replace them only if necessary.
CAUTION
Danger of pinching
Be careful when pressing the paper feed button not to get anything pinched in
the clamping mechanism and to avoid defects on the clamping mechanism.