
13
Maintenance
AVL 415SE Smoke Meter — Maintenance Guide
2.2.2
Service Engineer - Maintenance Schedule
For the disposal of environmentally sensitive substances, see the
AVL 415S
Smoke Meter Operating Manual
, chapter
Disposal of Environmentally Sensitive
Substances
.
Clamping lever
Annually
Visual check.
If damaged or if any part
is bent off.
Pump
Annually
Check and replace if necesary.
Every 2 000 - 3 000 oper-
ating hours.
Nipple & hose connec-
tions
Annually
Replace.
Annually
Cleaning of electronical
boards
Annually
Clean with high purity compressed
air.
If necessary or defective.
Filters of shop air filter
unit
Annually
Check and replace if necessary.
once every 1 … 2 years;
much more often for oil
lubricant compressors:
at least: every
1 … 2 months
Check valve
Annually
Check and replace if necesary.
Annually
Pressure reduction
assembly
Annually
Check.
After assembly.
Pressure switch of shop
air option
Annually
Check and adjust if necessary.
Annually
Component / Test
Lifetime /
Service interval
Maintenance
Replacement
Tab. 1
WARNING
Fire hazard
When using the required appliances like sprays, mounting pastes, lubricants,
always observe the relevant handling instructions and safety measures of the
various products.
All sprays used are a fire hazard and highly inflammable. Installation work using
the appliances on and inside measurement devices and on test beds must only
be carried out on cooled-down parts and in well ventilated areas.
NOTICE
The compressed air used to clean, has to be
absolutely
free from oil, water, and
particles.