background image

 

24-08-04 

AvaLight-HAL manual.doc 

 

 

 
Avantes 

www.avantes.com 

[email protected]

 

Lamp house 

Lamp socket 

Light bulb 

 

Adjustment of focusing in SMA fiber 

 

The AvaLight is factory adjusted to optimally focus the output into a 200µm fiber. If lower optical power is 
recommended or a different fiber (bundle) diameter is used, the optical power can be adjusted. 
1. 

Connect your fiber optic spectrometer or your optical power meter to a fiber to the AvaLight-HAL SMA 
socket (4) . 

2. 

Loosen blocking-screw  with delivered Allen key (1,3mm). 

3. 

By shifting the SMA socket you can optimize your optical power. 

4. 

Secure position by tightening the blocking screw . 

 

 

Changing the light source bulb 

 

1. 

Plug out the power connector from the socket. 

2. 

Remove screw protection caps on the frontside

 

3. 

Loosen 2 screws with philips screwdriver

 

4. 

turn the lightsource upside down

 

5. 

take out the frontplate and lamphouse

 

6. 

remove 2 nylon screws

 from lamp socket 

7. 

take out the lamp socket  

8. 

take out the bulb (carefull, may be hot) 

9. 

replace by new bulb, do not touch glass with 
your fingers 

10. 

put back lamp socket and nylon screws

 

11. 

Slide back front plate and lamp house, be 
carefull not to pinch the electrical wires

 

12.  put back screws and protection caps 

 
 
 
 

 

Отзывы: