background image

 

24-08-04 

AvaLight-HAL manual.doc 

 

 

 
Avantes 

www.avantes.com 

[email protected]

 

AvaLight-HAL Tungsten Halogen light source Operating Manual 

 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

 

 

 

PS-24VDC power supply 

Starting up 

1. 

Plug in the PS-24V power supply . 

2. 

Plug in the connector of the power supply into the socket of AvaLight-HAL. 

3. 

Connect the SMA-connectors of your fibers to the SMA-plug  

4. 

Switch on the Halogen lamp with the switch . 

 

Settings for high power or long lifetime  

 

The AvaLight-HAL is factory set for optimal power output (medium setting), with an internal jumper the optical 
energy can be controlled (see table).  
 

Jumper setting 

Optical output 

Color temperature 

Exp. Lifetime 

LOW 

70% 

Ca. 2.700K 

> 2.000 hrs. 

MEDIUM (default) 

100% 

Ca. 2.850K 

1.000 hrs 

HIGH 

150% 

Ca. 3.000K 

< 1.000 hrs. 

 
The internal jumper settings can be changed, please follow instructions: 

1. 

Remove screw protection caps on the backside

 

2. 

Loosen 2 screws with philips screwdriver

 

3. 

remove backplane

 

4. 

take out electronics board

 

5. 

Adjust jumper settings according to scheme below

 

6. 

put back electronics board, be carefull not to pinch the 
electrical wires

 

7. 

put back backplane, screws and screw protection caps 

 

 

 

 

SMA plug 

24V Power supply  

On-Off switch  

Screw caps  

Jumper JP1 

Отзывы: