24-08-04
AvaLight-CAL manual.doc
3
Avantes
www.avantes.com
HgAr Lamp replacement
1.
Plug out the power connector from the socket.
2.
Remove screw protection caps on the frontside
3.
Loosen 2 screws with philips screwdriver
4. take out the frontplate and electronics board
5. untighten 2 screws from HgAr socket
6. slightly bend the electrical wires out of HgAr
socket
7. take out the HgAr Lamp from Lamp holder
8. replace by new Lamp, make sure the lamp legs
make no short circuit.
9. Slide back electronics board and front plate,
tighten screws and put back screw
Technical Data
Power requirement
12VDC supply, 420 mA
Output
Hg and Ar lines from 254 nm to 922 nm
Optical power in 600µm fiber
1.6 mW
Connector
SMA 905 connector
Internal Voltage
1200 Volts AC at 30 kHz, 10 mA
Warm up
1 minute for vapor stabilization
Dimensions
175 x 110 x 44 mm
Ordering Information
AvaLight-CAL
Mercury-Argon
Spectral Calibration Source
AvaLight-CAL-RM
Rack mounted version of Mercury-Argon
Spectral Calibration Source
AvaLight-CAL-B
Replacement bulb, Mercury-Argon
PS-12V/1.25A
Power supply 100-240VAC/12VDC, 1.25A for AvaLight-CAL
HgAr socket
Lamp holder
Lamp holder
HgAr lamp