background image

 

24-08-04 

AvaLight-CAL manual.doc 

 

 

 
Avantes 

www.avantes.com 

[email protected]

 

 

HgAr Lamp replacement 

 

1. 

Plug out the power connector from the socket. 

2. 

Remove screw protection caps on the frontside

 

3. 

Loosen 2 screws with philips screwdriver

 

4. take out the frontplate and electronics  board  
5. untighten 2 screws  from HgAr socket 
6. slightly bend the electrical wires out of HgAr 

socket 

7. take out the HgAr Lamp from Lamp holder 
8. replace by new Lamp, make sure the lamp legs 

make no short circuit. 

9. Slide back electronics board and front plate, 

tighten screws and put back screw  

 

 
 
 
 
 
 
 
 
 
 
 
Technical Data 

Power requirement 

 

12VDC supply, 420 mA  

Output 

 

Hg and Ar lines from 254 nm to 922 nm  

Optical power in 600µm fiber 

1.6 mW  

Connector 

 

SMA 905 connector  

Internal Voltage 

 

1200 Volts AC at 30 kHz, 10 mA  

Warm up 

 

1 minute for vapor stabilization 

Dimensions 

 

175 x 110 x 44 mm  

 

Ordering Information 

AvaLight-CAL

 

Mercury-Argon

 

Spectral Calibration Source 

AvaLight-CAL-RM

  Rack mounted version of Mercury-Argon

 

Spectral Calibration Source 

AvaLight-CAL-B

 

Replacement bulb, Mercury-Argon

 

 

PS-12V/1.25A

 

Power supply 100-240VAC/12VDC, 1.25A for AvaLight-CAL 

 
 

HgAr socket 

Lamp holder 

Lamp holder 

HgAr lamp  