TROUBLESHOOTING MESH PROBLEMS
CLEANING THE ACCESSORIES
With time, the mesh holes may become blocked by
unknown particles. You may notice decrease output of
vapor. To test: Put 6 ml. of water inside the water tank
and power on the unit. If complete atomization time takes
longer than 30 minutes, then the mesh holes are clogged.
Solution A:
To clean the mesh holes and unclog the
blocked particles, put a solution of 2 drops of v 6
ml. of water into the water tank and allow the solution to
vaporize completely. Afterwards, rinse and dry thoroughly
before using it for a treatment.
Solution B:
Run the reverse self cleaning procedure with a
solution mix of 2 drops vinegar to 4 drops of water (refer
to the reverse self cleaning instructions),
If the problem persists after checking all possible causes,
please contact an authorized distributor to purchase a
new water tank.
Cleaning the accessories: Clean the accessories
(mouthpiece) with diluted soap water and rinse. Allow to
air dry completely
Cleaning Avya main unit: Clean the outside of Avya device
with a cotton cloth dampened with water. Do not allow
water to enter the device.
Cleaning the 4 stainless steel contact poles: Use a cotton
swab dampened with water to clean the stainless steel
contact electrodes on the Avya device (4 poles total - 2
poles on the main body and 2 poles underneath the water
tank).