If these solutions do not correct the problem, please contact your
distributor, or contact Aulisa by going online at
www.aulisa.com
under “Contact Us”.
This system is a precision electronic instrument andmust be repaired by
knowledgeable and specially trained Aulisa personnel only. Do not
attempt to open the case or repair the electronics.
Problem
Possible Solution
Check network usage/bandwith of other
interfaces within the Wi-Fi network that
may prevent the system from establishing
a connection.
Turn off the system, verify system
connections, and retry.
Ensure that the Oximeter Box is within
10 meters spherical radius to Receiver/
Transponder while being paired.
Make sure there are no headphones
inserted into the headphone jack of the
Display Unit.
The unit is in Alarm
mode, but no audible
alarms can be heard.
Cannot establish
system connection
Reset the Wi-Fi network for the
Display Unit.
Reset the Wi-Fi network for the
Receiver/Transponder.
Wait for two minutes and alarm tones will
automatically re-engage if it was silenced
temporarily.
52
Содержание Guardian Angel Rx GA2000
Страница 1: ...Guardian Angel Rx GA2000 Digital Vital Sign Monitoring System Instructions For Use TM...
Страница 11: ...Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable 8...
Страница 12: ...Stand Display Unit Charging Adaptor Oximeter Box Charging Adaptor Display Unit 9...
Страница 39: ...O X I M E T E R B O X B AT T E RY L O W John 36...
Страница 66: ...Service Support and Warranty Warranty under Contact Us 63...
Страница 68: ...65...
Страница 72: ...information aulisa com 69...
Страница 79: ...7MN00005 01...