6. Select the correct time zone.

    NOTE: On the top right of the MAIN screen of the Display Unit, ensure the 
                Receiver/Transponder connection indicator         is lit blue and the 
                Display Unit Wi-Fi connection indicator        has a strong signal.

Reset Wi-Fi Network

For Display Unit 

Use the following procedures to change the Wi-Fi network of the
Display Unit:

1. On the Display Unit, tap on the “SETTINGS” icon. Then, tap on 
    “CONNECTION”.

2. Tap on “DISPLAY UNIT”.

3. Select the desired Wi-Fi network from the list of available connections 
    displayed on the MAIN screen. Enter the network security password. 
    Then, press “CONFIRM”.

For Receiver/ Transponder 

Use the following procedures to reset the Receiver/Transponder to link to 
the desired Wi-Fi network:

NOTE: The Wi-Fi network of the Display Unit will be changed, together
            with that of the Receiver/Transponder.

1. On the Display Unit, tap on the “SETTINGS” icon, then tap on
    “CONNECTION”.

2. Tap on “RECEIVER/TRANSPONDER”. Then, tap on “CONFIRM”.

26

Содержание Guardian Angel Rx GA2000

Страница 1: ...Guardian Angel Rx GA2000 Digital Vital Sign Monitoring System Instructions For Use TM...

Страница 2: ...manual shall imply Taiwan Aulisa Medical Devices Technologies Inc Aulisa is a registered trademark of Taiwan Aulisa Medical Devices Technologies Inc Read this entire manual carefully before using Aul...

Страница 3: ...rging 34 44 Alarms Charging the Oximeter Box 34 44 Powering the Display Unit 38 46 Alarm Delay Function Powering the Receiver Transponder 42 47 43 Care and Maintenance 48 Troubleshooting 49 Technical...

Страница 4: ...o instruction manual Indicates separate collection for electrical and electronic equipment WEEE Non ionizing electromagnetic radiation Equipment includes RF transmitters Interference may occur in the...

Страница 5: ...Prescription use only Temperature limit NON STERILE Non sterile Date of Manufacturer Classification for water ingress and particulate matter IP22 Warning Charging Port 2...

Страница 6: ...les are manufactured to meet the accuracy specifications for Aulisa GA2000 Digital Vital Sign Monitoring System Using other manufacturers sensor cables can result in improper pulse oximeter performanc...

Страница 7: ...arterial oxygen saturation of functional hemoglobin Significant levels of dysfunctional hemoglobin such as methemoglobin might affect the accuracy of the measurement 13 Anemia may affect the accuracy...

Страница 8: ...Sign Monitoring System fails to respond as described discontinue use until the situation is corrected by qualified personnel 4 Cardiogreen and other intravascular dyes may affect the accuracy of SpO2...

Страница 9: ...ng instructions regarding disposal or recycling of the device and device components including batteries 16 Do not subject the system to extreme hot or cold temperatures humidity or direct sunlight 17...

Страница 10: ...e Aulisa GA2000 Digital Vital Sign Monitoring System hereinafter referred to as Aulisa GA2000 system The system includes the following components and accessories The system comes with either two Pedia...

Страница 11: ...Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable 8...

Страница 12: ...Stand Display Unit Charging Adaptor Oximeter Box Charging Adaptor Display Unit 9...

Страница 13: ...Sticker Receiver Transponder Clamp Receiver Transponder Velcro Strap Receiver Transponder Charging Adaptor Receiver Transponder 10...

Страница 14: ...eusable device Principle of Operation Aulisa GA2000 Digital Vital Sign Monitoring System measures SpO2 and pulse rate based on non invasive light emitting diode LED transmittance technology measuring...

Страница 15: ...nt for vital sign monitoring It features a sensor chip along with electronics for vital sign measuring and analyzing The Oximeter Box must be used within 10 meters from the Receiver Transponder 1 2 3...

Страница 16: ...data from the Oximeter Box via Bluetooth integrates audio and video of the patient and then converts the data to Wi Fi signals which are transmitted to and displayed by the Display Unit 1 2 4 5 6 Pow...

Страница 17: ...beep made audible upon an alarm condition trigger event The Display Unit incorporates a talking and listening function that allows audio messages to be received and sent via the Receiver Transponder...

Страница 18: ...s the window showing the pulse amplitude Pulse Amplitude When the vital signs cannot be measured the Display Unit shows dashes in each of the vital sign windows No data SpO2 This icon displays whether...

Страница 19: ...m alarm will be displayed on the Display Unit when the Oximeter Box battery is low Battery Level of Oximeter Box These icons signify the battery level of the Display Unit A medium priority system alar...

Страница 20: ...identifies an alarm condition exists represents high priority and represents medium priority Alarm Indicator This icon indicates that the alarm is turned off for the corresponding physiological condit...

Страница 21: ...e MAIN screen to access the settings menu of the system Return to Previous Screen Tap on this button on the top right corner of the screen to return to the previous page Alarm History Tap on this butt...

Страница 22: ...t Time Zone In the settings menu tap on this button to select the correct time zone for the Display Unit Connection Setup In the settings menu tap on this button to change the Wi Fi network of the Dis...

Страница 23: ...riggered Tap on the button to temporarily silence the alarm audio of the current triggered alarm event for 2 minutes Turn Off Alarm Audio The button appears on the MAIN screen when an alarm is trigger...

Страница 24: ...ns below to assemble the Receiver Transponder NOTE Securing the power cable to the holder with the Velcro strap can prevent the Receiver Transponder body from falling down when it accidentally falls o...

Страница 25: ...the Display Unit 5 Follow the onscreen instructions on the Display Unit to complete the system setup See System Connection section on page 25 for more information NOTE On the top right of the MAIN scr...

Страница 26: ...The wristband should be worn with the arrow indicator facing towards the patient s hand 8 Insert the Oximeter Box into the holder Attach the sensor probe to the thumb or finger making sure that the s...

Страница 27: ...Oximeter Box under the control of the Display Unit NOTE The Oximeter Box must be placed within 10 meters 32 8 feet from the Receiver Transponder NOTE The Bluetooth connection status on the Display Un...

Страница 28: ...iver Transponder plays a melody around a minute 4 Select a secure customer Wi Fi network from the list of available connections displayed on the Display Unit Enter the network security password Then p...

Страница 29: ...SETTINGS icon Then tap on CONNECTION 2 Tap on DISPLAY UNIT 3 Select the desired Wi Fi network from the list of available connections displayed on the MAIN screen Enter the network security password T...

Страница 30: ...ovides no security and exposes al your network traffic NOTE Ensure the customer Wi Fi network connection has a strong and reliable signal 5 Wait until the setting up of the Wi Fi connection for the Di...

Страница 31: ...x NOTE Up to four Oximeter Boxes can be stored on the Display Unit 1 Ensure the desired Oximeter Box is turned on See Power and Charging section on page 44 for more information NOTE The power On Off L...

Страница 32: ...olding the power On Off button for 3 seconds Wait until the Wi Fi Link LED lights red See Power and Charging section on page 47 for more information NOTE The power On Off LED on Receiver Transponder l...

Страница 33: ...if an alarm message and audio signal is not generated 5 Press on the PAUSE AUDIO button to temporarily silence for 2 minutes 6 After alarm signal is regenerated press on the AUDIO OFF button to silenc...

Страница 34: ...the Receiver Transponder connection indicator is lit blue and the Display Unit Wi Fi connection indicator has a strong signal 5 Verify that a SpO2 reading is displayed that a pulse rate value appears...

Страница 35: ...o Talking Press and hold to send an audio message to patient 2 3 NOTE When patient privacy is desired follow the instructions below to place the sticker over the camera and microphone of Receiver Tran...

Страница 36: ...on on the MAIN screen Oximeter Box Click the power On Off button to turn off the Oximeter Box NOTE When the power is turned off the power On Off LED lit green will turn off NOTE The Oximeter Box will...

Страница 37: ...re those that require immediate attention to the patient They include SpO2 and pulse rate alarms On the Display Unit high priority alarms are indicated with rapid blinking vital sign readings in red c...

Страница 38: ...y Unit NOTE The following table describes alarm conditions and visual indicators Alarm Condition Medium Priority Alarm Visual Indicator Sensor Probe Detached from Patient Oximeter Box Battery Low Sens...

Страница 39: ...O X I M E T E R B O X B AT T E RY L O W John 36...

Страница 40: ...exists if different alarm presets are used for the same or similar equipment in any single area Do not plug a headphone into headphone jack of the Display Unit as this will significantly reduce the v...

Страница 41: ...tem connection is established See System Connection section on page 25 more information 2 Tap on SETTINGS button on the MAIN screen and then tap on ALARM LIMITS button NOTE Alarm limits can be adjuste...

Страница 42: ...ff tap on ON OFF button Turn on the alarm before adjusting the value NOTE SpO2 max limit is turned off by default NOTE There is no alarm setting for pulse amplitude John PR bpm 150 50 40 PULSE RATE LO...

Страница 43: ...s or drag the seekbar to adjust the values NOTE The minimum alarm limit cannot exceed the max alarm limit even if the max alarm limit is turned off For example if the max SpO2 limit is turned off but...

Страница 44: ...ned off the alarms will no longer be displayed or sound Follow the instructions above to turn on the alarms Consult a physician about the appropriate vital signlimits for the user before adjusting an...

Страница 45: ...lowing table lists the alarm triggering delay statistics for SpO2 Low Alarm Limit SpO2 Low Alarm Limit Maximum Delay seconds Average Delay seconds 74 20 7 6 50 16 6 4 5 4 90 86 90 81 85 76 80 75 Likew...

Страница 46: ...ttings can be restored only when the system connection is established 2 Tap on SETTINGS button located on the top right corner of the MAIN screen 3 Tap on DEFAULT ALARM button 4 Tap on CONFIRM button...

Страница 47: ...wing the steps below NOTE The Display Unit will alarm the user when the Oximeter Box is low on battery Once on low battery the Oximeter Box will work for up to another 2 hours working time on low batt...

Страница 48: ...ge the Oximeter Box NOTE Verify operation of the system See Verifying System Operation section on page 30 for more information and check the battery status on the MAIN screen of the Display Unit NOTE...

Страница 49: ...e and power the Display Unit The Display Unit will alarm the user when the Display Unit itself is low on battery 1 Plug the Type C end of the charging adaptor black with a Type C end into the Display...

Страница 50: ...power outlet NOTE The AC power LED indicator will light green when AC adaptor is ON NOTE The charging LED indicator will light blue when internal battery is charging and will turn off when it is full...

Страница 51: ...quids to enter any openings in the device Allow the unit to dry thoroughly before reuse Do not immerse the device in liquid and do not use caustic or abrasive cleaning agents on the device The advance...

Страница 52: ...nnected to the outlet and the Display Unit Reposition the sensor probe or reinsert the finger and keep the hand motionless for at least ten 10 seconds Position the sensor probe at a different site All...

Страница 53: ...e sure the sensor probe is attached to the finger securely Make sure the sensor cable is securely attached to the Oximeter Box Check the Display Unit for any alarms or error messages Check the sensor...

Страница 54: ...system check system connections and retry Reposition the sensor probe or reinsert the finger and keep the hand motionless for at least ten 10 seconds Insert the sensor probe to a finger without finge...

Страница 55: ...Wi Fi network that may prevent the system from establishing a connection Turn off the system verify system connections and retry Ensure that the Oximeter Box is within 10 meters spherical radius to R...

Страница 56: ...s of measurement with multiple observations per individual Journal of Biopharmaceutical Statistics 17 571 582 RMS Error SpO SaO 2 2 2 n Root mean square error RMS error is calculated as follows 1Bias...

Страница 57: ...e accuracy has been functionally tested against an electronic pulse simulator at 30 50 80 100 150 200 250 and 290 bpm with combinations of Pulse Amplitude settings of 0 5 1 3 5 7 10 13 15 17 and 20 an...

Страница 58: ...for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment Guidance and Manufacturer s Declaration Electromagnetic E...

Страница 59: ...ation Electromagnetic Immunity Immunity test Electromagnetic environment guidance 8 kV contact 15 kV air Electrostatic Discharge ESD IEC 61000 4 2 Electrical Fast Transient Burst IEC 61000 4 4 Surge I...

Страница 60: ...4 11 Power Frequency 50 60 Hz Magnetic Field IEC 61000 4 8 30 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environm...

Страница 61: ...cified below The customer and or user of this device should ensure that it is used in such an environment Guidance and Manufacturer s Declaration Electromagnetic Immunity Immunity test Electromagnetic...

Страница 62: ...h in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional m...

Страница 63: ...the separation between the equipment and receiver 3 Connect the equipment into an outlet on a circuit different from that to which the receiver is connected 4 Consult the dealer or an experienced rad...

Страница 64: ...operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful inte...

Страница 65: ...range of 5 15 to 5 25GHz to reduce the potential for harmful interference to co channel mobile satellite systems High power radar is allocated as the primary user of the 5 25 to 5 35GHz and 5 65 to 5...

Страница 66: ...Service Support and Warranty Warranty under Contact Us 63...

Страница 67: ...n Personal Information and we will protect it WE AT AULISA VALUE KEEPING YOUR PERSONAL INFORMATION CONFIDENTIAL AND USING IT SOLELY IN THE CONTEXT OF OUR MISSION TO PROVIDE CONTINUOUS MONITORING OF VI...

Страница 68: ...65...

Страница 69: ...its user demographics and behavior based on the Personal Information we collect from you and the other information provided to us This research may be compiled and analyzed on an aggregate basis and...

Страница 70: ...Aulisa Related Companies for purposes consistent with this Privacy Policy With Our Agents Consultants and Related Third Parties Aulisa like many businesses sometimes hires other companies to perform...

Страница 71: ...easonable physical technical and administrative security controls to maintain the security and integrity of your Personal Information to protect against any anticipated threats or hazards to the secur...

Страница 72: ...information aulisa com 69...

Страница 73: ...n the Device Information Submission by Minors If the Device is being utilized by a minor and the Devices are being used to monitor a minor you represent warrant and covenant that by agreeing to the te...

Страница 74: ...se Rate 1 to 100 30 to 290 bpm 70 100 3 digits 3 Specifications Alarms SpO2 Default SpO2 Limit Upper Limit Lower Limit Alarm Limit Range Upper Limit Lower Limit Adjustment Step Pulse Rate Default Puls...

Страница 75: ...11 b g n ac 2 4GHz 5GHz Bi direction Temperature Operating Storage Transportation 5 C to 40 C 25 C to 70 C Operating Altitude Operating Storage Transportation 15 to 90 non condensing 10 to 93 relative...

Страница 76: ...Input 100 240 V AC 50 60 Hz 5 V DC AC adaptor Internal Power Battery Operating Life Dimensions Weight Wireless Communication Alarm Sound Pressure Ingress Protection 3 8 V battery 2 hours of continuou...

Страница 77: ...imensions Body Stand Weight Body Stand Wireless Communication Ingress Protection 2 6 x 4 6 x 2 7 67 mm x 116 mm x 69 mm 3 5 x 3 6 x 1 3 90 mm x 91 mm x 32 mm 6 oz 171g 1 oz 33 g Bluetooth Wi Fi IP22 7...

Страница 78: ...Charging Adaptor Oximeter Box Charging Adaptor Display Unit Charging Adaptor Receiver Transponder For more information about Aulisa parts and accessories contact your distributor or contact Aulisa by...

Страница 79: ...7MN00005 01...

Отзывы: