43
The graph below shows the error (SpO – SaO ) plots of each subject
2
2
measured by the Aulisa GA1000 with upper and lower 95% limits of
agreement. Each sample data point is from a clinical study in healthy
adult volunteers.
Pulse Rate Accuracy
Pulse rate accuracy has been functionally tested against an electronic
pulse simulator at 30, 50, 80, 100, 150, 200, 250, and 290 bpm, with
combinations of pulse amplitude settings of 0.5, 1, 3, 5, 7, 10, 13, 15,
17 and 20, and SpO settings of 100%, 90%, 80%, 70%, 60%, 50%,
2
40%, 30%, 20%, 10%, and 1%. All 880 combinations of testing points
(=8 x 10 x 11) of pulse rate passed the ± 3% acceptance criteria.
Equipment Response Time
Aulisa GA1000 Digital Vital Sign Monitoring System uses a moving
average to determine the pulse rate and SpO . The following table
2
shows the equipment response time of Aulisa GA1000 Digital Vital Sign
Monitoring System.
Equipment
DelaysDelay(Seconds)
Data Averaging
≤ 4 seconds
Alarm Condition Delay
≤ 4 seconds
Alarm Signal Generation Delay
0 seconds
Data Update Period
1 second
2
Содержание Guardian Angel Rx GA1000
Страница 1: ...Guardian Angel GA1000 Digital Vital Sign Monitoring System TM Rx...
Страница 11: ...8 Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable...
Страница 12: ...9 Charging Adaptor Oximeter Box Stand Display Unit Charging Adaptor Display Unit...
Страница 33: ...30 In an alarm event ALARM LIMITS button will appear after you select AUDIO PAUSE button or AUDIO OFF button...
Страница 69: ...7MN00001 02...