32
4. Tap on ʻʻ + ʼʼ or ʻʻ– ʼʼ buttons or drag the ʻʻseekbarʼʼ to adjust the values.
NOTE:The minimum alarm limit cannot exceed the max alarm limit,
even if the max alarm limit is turned off. For example, if the
max SpO limit is turned off but was previously set at 90%,
2
the min SpO limit cannot be set higher than 90%. If you
2
want to set min SpO limit at 90%, turn on the max SpO
2
2
limit, set it above 90% and turn it off again as you wish.
NOTE:The following table describes the default settings,
adjustment ranges, and intervals.
High Priority Alarm
SpO Upper Alarm Limit
2
SpO Lower Alarm Limit
2
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Factory
Default
Adjustment
Interval
Off
85%
150 bpm
50 bpm
Off, 85 to 100
Off, 50 to 95
Off, 75 to 275
Off, 30 to 110
Adjustment
Options
1% SpO
2
1% SpO
2
1 bpm
1 bpm
5. Tap on ʻʻCONFIRMʼʼ to save the alarm limits.
NOTE:The SpO and pulse rate upper and lower alarm limits
2
appear as smaller values to the top right and bottom right of
their respective window on the MAIN screen.
.
The new ALARM LIMITS do not go into effect until the “CONFIRM” button
is tapped.
Содержание Guardian Angel Rx GA1000
Страница 1: ...Guardian Angel GA1000 Digital Vital Sign Monitoring System TM Rx...
Страница 11: ...8 Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable...
Страница 12: ...9 Charging Adaptor Oximeter Box Stand Display Unit Charging Adaptor Display Unit...
Страница 33: ...30 In an alarm event ALARM LIMITS button will appear after you select AUDIO PAUSE button or AUDIO OFF button...
Страница 69: ...7MN00001 02...