Check the Display Unit for any alarms or
error messages.
Check the Oximeter Box for power.
Check the sensor cable for any visible
signs of deterioration.
Shield the sensor probe from the light
source.
Insert the sensor probe to a finger without
fingernail polish or an artificial nail.
Make sure the sensor probe is attached to
the finger securely.
Check the sensor cable for any visible signs
of deterioration.
Reduce patient motion.
Make sure the sensor cable is attached to
the finger securely.
Make sure the Oximeter Box is within
10 meters (32.8 feet) spherical radius to the
Display Unit and that the devices are paired.
Unstable/Constant
SpO and Pulse
2
Rate readings
Reposition the sensor probe or
reinsert the finger and keep the hand
motionless for at least 10 seconds.
Position the sensor probe at a different site.
Make sure the sensor cable is securely
attached to the Oximeter Box.
Make sure the sensor cable is securely
attached to the Oximeter Box.
Position the sensor probe at a different site.
Verify that the Oximeter Box is paired with
the Display Unit.
Turn off the unit, check all connections, and
retry.
Position the sensor probe at a different site.
Data update period
has exceeded limit
Problem
Possible Solution
Содержание Guardian Angel Rx GA1000
Страница 1: ...Guardian Angel GA1000 Digital Vital Sign Monitoring System TM Rx...
Страница 11: ...8 Adult Oximeter Sensor Cable Wristband Pediatric Oximeter Sensor Cable...
Страница 12: ...9 Charging Adaptor Oximeter Box Stand Display Unit Charging Adaptor Display Unit...
Страница 33: ...30 In an alarm event ALARM LIMITS button will appear after you select AUDIO PAUSE button or AUDIO OFF button...
Страница 69: ...7MN00001 02...