![AUDAC LCR700 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/audac/lcr700/lcr700_user-manual-and-installation-manual_3006949009.webp)
8
Specifications
CAPACITANCE
:
Test Frequency 120 Hz
Range
Min.
Max.
Cx
DF
Note
20mF
1µF
10.000mF
+/- (5.0%rdg +
5counts)
DF<0.1
+/- (10%rdg +
100/Cx + 5counts)
DF< 0.1
After short cal.
2000µF 100nF 1999.9µ
+/- (1.0%rdg + 5
counts)
DF<0.1
+/- (2.0%rdg +
100/Cx + 5counts)
DF<0.1
After short cal.
200µF
10nF
199.99µF
+/- (0.7%rdg + 3
counts) DF<0.5
+/- (0.7%rdg +
100/Cx + 5 counts)
DF<0.5
-
20µF
1nF
19.999µF
+/- (0.7%rdg +
3counts) DF<0.5
+/- (0.7%rdg +
100/Cx + 5counts)
DF<0.5
-
2000nF 100pF 1999.9nF
+/- (0.7%rdg +3counts)
DF<0.5
+/- (0.7%rdg +
100/Cx + 5counts)
DF<0.5
-
200nF
10 pF
199.99nF
+/- (0.7%rdg + 5
counts) DF<0.5
+/- (0.7%rdg +
100/Cx + 5counts)
DF<0.5
After open cal.
20nF
1pF
19.999nF
+/- (1.0%rdg +
5counts) DF<0.1
+/- (2.0%rdg +
100/Cx + 5 counts)
DF<0.1
After open cal.
Chapter
3