16
7. REL setting:
Press the “RANGE” button to select the desired range. Press the “SET” button to enter the set mode and then
the “REL” button. The “
” symbol and the left digit of the 4 ½ digits will now flash for setting a reference
value. After entering a number the next digit will flash. When the reference value is set, press “ENTER” to
return to the normal mode.
10. REL Relative mode (only Main display)
Press the “REL” key to enter the Relative mode. The displayed reading is stored as a reference value and the
display will indicate zero or a value close to zero (since the measured value and the reference value are the same
at this point). Indicator “
” is displayed. Press the “REL” key again to exit the Relative mode.
For example: the displayed reading is 100.0, then press REL to store as a reference value and the display
indicates zero. If you have set 100.0 as a reference value and if the input signal is 99.5, the relative mode will be
-0.5. (99.5-100.0).
11. Hi/Lo Limits
Press “Hi/Lo LIMITS” to enter the “Hi/Lo LIMITS” mode, and change to manual range mode. The original
“Hi/Lo LIMITS” values and indicators “
” and “
”appear individually and at the same time. When the
input exceeds Hi Limit, the “
” indicator is blinking and emits a continuous tone. When the input is lower
than the Lo Limit, the “
” indicator is blinking and a beep tone is heard. Press “Hi/Lo LIMITS” again to exit
this mode. In addition, the meter ignores and does not record if an overload (“OL”) situation occurs in the
comparative processes or in the capacitance range <=50 counts.
12. TOL
Press “TOL” to enter the tolerance mode, and change to manual range mode automatically. The original preset
standard value and the “TOL” indicator appear individually and at the same time. When you enter the TOL
mode, the Main display indicates the reference value and the second display indicates the tolerance value.
There are 4 preset values in the TOL mode: just press TOL again to step through 1 %, 5 %, 10 % and 20%.
The meter will emit an audible tone and blink the upper “
”or lower “
”indicator if the measured value is
outside the limits. Standard values can be set by the “SET” function. In addition, the meter ignores and does
not record if an overload (“OL”) situation occurs in the comparative processes or in the capacitance range <=
50 counts. Press “TOL” for 2 seconds to exit this function.