13.Data transmission settings
You can change the settings and the output data via USB.
■
Data Transmission Protocol
Stop bit: 1 bit
Parity bit: None
Delimiter: CR + LF
Flow control: None
■
Transmission and reception code list
"Command"
Contents
Cautions
Example (all with CR+LF attached at the end)
The signal is transmitted at 2KHz.
(The output continues until the "STOP" signal is received.)
"STOP"
Continuous transmission stop
STOP
"MEMORY"
Transmit all stored data
MEMORY
"CLEAR"
ALL CLEAR
Delete the stored data
CLEAR
"FUNC_MODE"
FUNC Switching
Please send first when you change the settings.
FUNC_MODE
"TRACK"
Change to TRACK
TRACK
"PEAK"
Change to PEAK
PEAK
"PP"
Change to PP
PP
"GF"
Change to gram unit system
GF
"NEWTON"
Change to Newton unit system
NEWTON
"POND"
Change to pound unit system
POND
Error if more than the settable value.
Cannot exceed the maximum value.
Error if more than the settable value.
Cannot exceed the comparator HI setting.
"OVER:"
Overload setting value
Error if not a settable value
.
10
~
120 in 10 increments
OVER:100
"LCD_LATE:"
Screen update rate
Error if not a settable value
.
20,10,5,2 only
LCD_LATE:2
"LCD_FLIPPED:"
Flip the screen vertically
"ON": permission, "OFF": none
LCD_FLIPPED:ON
"POWER_OFF:"
Auto power off
"ON": permission, "OFF": none
POWER_OFF:ON
"AUTO_CLEAR:"
ALL CLEAR
"ON": permission, "OFF": none
AUTO_CLEAR:ON
"MIN_DIGIT:"
Minimum digit display
"ON": permission, "OFF": none
MIN_DIGIT:OFF
"BUZZ:"
Buzzer permission
"ON": permission, "OFF": none
BUZZ:OFF
"BACK_LIGHT:"
Backlight permission
"ON": permission, "OFF": none
BACK_LIGHT:ON
"GAIN_INVERT:"
Plus / minus inversion
"ON": permission, "OFF": none
GAIN_INVERT:OFF
"INIT_VALUE:"
Reset to initial setting
"ON": permission, "OFF": none
INIT_VALUE:
"PRINTER:"
Printer permission
"ON": permission, "OFF": none
PRINTER:ON
"SET_SAVE"
Save change value
Send it at last to save values.
SET_SAVE
"CANCEL"
Cancel the change
Send it at last to cancel the change.
CANCEL
"VALUE_READ"
Read current value
Read current value. COM_OK is not returned.
VALUE_READ
* Receive a reply message of "COM_OK!!" if the command input is correct.
* Receive a reply message of "COM_ERR!!" if the command input is incorrect.
* Be sure to send "SET_SAVE" after changing the settings. Otherwise, the settings will not be enabled and the display
will not return to the measurement screen, either.
* Once set, it cannot be returned to the original value.
* An example of send command input
Set the comparator HI to 10.0.
Command: setting value CR + LF
COMP_HI: 10.0 CR+LF
"STAT"
"CMP_HI:"
"CMP_LO:"
Continuous transmission
Comparator HI setting
Comparator LO setting
STAT
CMP_HI:2000
CMP_LO:1000
Baud rate: 19200
Character length: 8 bits