Service Manual
Auto Setup
ADS1000 Series Digital Storage Oscilloscopes have auto setup function, which can identify the
type of waveform and adjust controls to produce a usable display of the input signal.
AUTO is the auto setup button.
Table2-1 Auto setup function menu:
Option
Introduction
(
Multi-cycle
sine
)
Auto set the screen and display several cycle
signal.
(
Single-cycle
sine
)
Set the screen and auto display single cycle signal.
(
Rising
edge
)
Auto set and show the rising time.
(
Falling
edge
)
Auto set and show the falling time.
(
Undo
Setup
)
Causes the oscilloscope to recall the previous
setup.
There are four measure parameter will be displayed in the screen. Auto set determines the trigger
source based on the following conditions:
●
If multiple channels have signals, channel with the lowest frequency signal.
●
The input frequency must be higher than 10Hz, and the amplitude must be higher than
8MV
Table 2-2 Auto setup function item
Function
Function
Function
Function
Setting
Setting
Setting
Setting
Acquire Mode
Adjusted to Sampling
Display Format
Y-T
Display Type
Set to Dots for a video signal, set to Vectors
for an FFT spectrum; otherwise, unchanged
Vertical Coupling
Adjusted to DC or AC according to the input
signal