6

2.0    Safety advice

 

 The 

Variotherm plus

 unit is produced according to 

  IEC 601/EN 60601 and listed in the following classes:

  •  VDE Class of protection 1

  •  Class IIa (EEC 93/42).

  The appliance may be installed and operated only in  

  examination units from ATMOS.

 The 

Variotherm plus

 may be used in supervised operation  

  by  qualified  personnel  only  who  have  been  authorised 

 

  by ATMOS and who have been trained for operating the  

  appliance (IEC 601-1/EN 60601-1).

  The mains voltage specified on the type plate must  

  correspond with the data of the power supply system.

  Prior to every application of the equipment make sure that  

  it is technically safe and in proper condition. Damaged  

  leads must be replaced immediately!

  Correct configuration in assembly of country-specific 

   connections

  green/yellow: 

protective conductor (PE)

  blue: 

neutral conductor (N)

  black resp. brown:  phase (L)

  The control panel must be well visible to and in reach of  

  the operator. 

  Metal parts can be hot!

  Prior to spraying, the water temperature must be checked 

  by the user!

  Switch off main switch of the examination unit after  

  finishing work in practice and 

close water cock of water  

 supply.

  The ATMOS Variotherm may be operated only in

  rooms used for medical purposes, but not in areas

  subject to explosion hazards and in oxygen rich
 environments.

 

Additional equipment, which is to be connected to the ana-

log and digital interfaces of the appliance, must provably 

meet the requirements of relevant EN specifications (for 

inst. EN 60950 for data processing equipment and EN 

60601 for electrical medical appliances). Furthermore, all 

configurations must satisfy system specification EN 60601-

1-1. Whoever connects additional equipment to the signal 

input or signal output section is a system configuration ope

-

rator and as such responsible for meeting the requirements 

of system specification EN 60601-1-1. In case of further 

questions please contact your local specialist supplier or 

the technical service.

 

The irrigation jet must not get into contact with conta- 

  minated material.

 

Application of the warm water jet only with tube tip 

  put on!

 

Caution to be paid to avoid injury to eardrum when  

  introducing the jet!

 

For hygienic reasons jet connections must be changed  

  after each patient. This prevents also retrograde ger- 

  mination of the warm water system.

 

Use only for irrigation of auditory canal !

  ATMOS is not liable for personal injury and damage to  

  property if

  •  no original ATMOS parts are being used,

  •  the advice for use in these operating instructions is not

 

    being observed,

  •  assembly,  new  settings,  alterations,  extensions  and 

 

    repairs have been carried out by personnel not  

 

  authorised by ATMOS.

 

Prior to ear rinsing, the user has to control the output  

  pressure, it may not exceed 2.2 bar.

 

Please note:

  A medical insulating transformer with earth leakage 

  monitor or any similar safety system acc. to  

  EN 60 601-1 is required, if several devices are  

  connected over one common power supply. The  

  transformer must correspond to the power  

  consumption of all the devices to be connected.

 

 The flow rate has to be checked after switch on or 

 

  at least once a day with a measuring cup. The quantity  

  of 500 ml/min may not be exceeded. The water jet has  

  to be straight.

Содержание Variotherm

Страница 1: ...s English ATMOS MedizinTechnik GmbH Co KG Ludwig Kegel Str 16 79853 Lenzkirch Germany Tel 49 0 7653 6 89 0 Fax 49 0 7653 6 89 190 atmos atmosmed de www atmosmed de MedizinTechnik 2013 08 Index 16 ATMOS Variotherm 530 2030 B ...

Страница 2: ...f the Variotherm 12 4 2 Temperature adjustment 12 4 3 Time setting 12 4 4 Thermal nystagmus stimulation 12 5 0 Cleaning 5 1 General information on cleaning and disinfection 13 5 2 Recommended disinfectants 14 5 4 Cleaning method for handle with jet connection for water irrigation tip 14 6 0 Maintenance and service 6 1 Functional test 15 6 2 Visual inspection 16 6 3 Electrical test 16 6 4 Cleaning ...

Страница 3: ...d careful work provides best protection against accidents Operational safety and readiness for use depend not only on your capabilities but also on care and maintenance given to the Variotherm plus For this reason regular cleaning and service work are a must Major maintenance and repair work may be carried out only by expert personnel authorised by ATMOS In case of repairs you should insist that o...

Страница 4: ...ation of the vestibular organ Application For the application on humans Specifications of the main functions Produces a jet of water for the ear irrigation at body temperature 37 C with a flow rate of 400 ml min for the removal of cerumen Stimulation of the vestibular organ with a defined flow and a fixed temperature for the functional testing Application organ Auditory canal to the drumhead Durat...

Страница 5: ... Timer adjustment in seconds Timer Cold stimulation 20 36 C Warm stimulation 38 47 C Rinsing mode water temperature 37 C C s Low High Reduced flow for stimulation of the vestibule High flow for rinsing the auditory canal Setpoint value adjustment possible Increase of setpoint value Decrease of setpoint value 1 0 Introduction Handle ...

Страница 6: ...cessing equipment and EN 60601 for electrical medical appliances Furthermore all configurations must satisfy system specification EN 60601 1 1 Whoever connects additional equipment to the signal input or signal output section is a system configuration ope rator and as such responsible for meeting the requirements of system specification EN 60601 1 1 In case of further questions please contact your...

Страница 7: ... is hygienically carried out Before screwing them into place disinfect threaded connections with 70 alcohol When carrying out maintenance work on parts which come into contact with the water wear disposable gloves Carry out thermal cleaning at least 3 times before commissioning This work is carried out by the ATMOS service technician Running operation Exchange and disinfection required for contact...

Страница 8: ...uantity 1 Variotherm basic unit 1 Irrigation handle with Double barrelled hose 3 Jet connection short straight 80 mm 1 Splash protection 1 Rubber bush for splash protection 1 Three holes gasket 1 Hose tips for jet connection 30 pcs 1 Operating instructions ...

Страница 9: ...tion time display double digits resolution 1s Temperature display double digits resolution 1 C shows actual value Start stop key for stimulation 150 ml min Adjusting key for stimulation time Key for irrigation mode 37 C Key for warm stimulation e g 44 C Key for cold stimulation e g 30 C 3 0 Installation and start up ...

Страница 10: ...y Prior to stimulation turn the set screw fig 2 upwards and arrest it 2 adjusting possibilities High Flow quantity for rinsing the auditory canal Low Flow quantity for stimulating the vestibular organ Prior to ear irrigation the set screw is turned downwards inserted into the handle and then adjusted by turning it downwards again For fine adjustment of the flow the adjusting screw fig 2 is turned ...

Страница 11: ...oles gaskets as otherwise correct function of the unit is not guaranteed Insert a three holes gasket in the irrigation handle and screw it on the free end of the double barrelled hose Open the water tap of the ENT examination unit Check whether all connections are tight 3 4 Start up Insert the handle in its holder the jet must point to reverse side of the unit Switch on the examination unit Automa...

Страница 12: ... previous time setting is taken over automatically Please check the time setting of the timer once in a while 4 2 Temperature adjustment 4 4 Thermal nystagmus stimulation Fig 3 irrigation handle set screw adjusting screw Remove the handle select and if need be adjust the temperature s 4 3 if required adjust the time press is blinking approx 10 sec as soon as the Variotherm is ready a beep is audib...

Страница 13: ...cause stress cracks in the plastic material used for the housing of the unit For cleaning and disinfection you may also use disinfectant sprays or disinfectant tissues Switch off the ENT unit by the main switch before starting cleaning and disinfection Wipe the unit surface with a cloth moistened with a cleaning or disinfecting solution Make sure that no liquid penetrates the unit You may use all ...

Страница 14: ... Weigert Hamburg forte non ionic tensides 5g Application concentrate enzymes preserving agents Machine disinfection Ingredients in 100 g Manufacturer Dismozon pur magnesium monoperoxyphthalate Bode Chemie Hamburg Application concentrate hexahydrate 80 g Green Clean SK alkyl dimethyl benzyl ammonium chloride 1 g Metasys Rum Austria Application concentrate dialkyl dimethyl ammonium chloride Alkyldim...

Страница 15: ...ignal during the first second after switching on in particular segmental test 8 8 Set all temperature setting switches on 40 C and test consecutively by touching the corresponding keys whether in all ranges 40 C is reached in ACTUAL temperature indication Spray off and check water temperature by hand Carry out the same test with temperatures 37 C and 28 C Measure temperature of water splashed off ...

Страница 16: ...r inst Citrosteril by Fresenius Screw glass on again Open water tap again Switch on the unit Decalcification now takes place automatically in the thermal process for reducing bacterial count If after switching on the unit has been in the thermal process for reducing bacterial count for 5 minutes the water passages have been decalcified and cleaned and also sufficiently flushed The Variotherm plus ...

Страница 17: ...k whether temperature setting is too high If necessary adjust desired temperature to a value of 48 C by means of the coding switches Inform service staff E7 temperature too high 48 C display only in the stimulation and irrigation mode but not in the thermal process for reducing bacterial count Error in temperature display Have temperature feeler of the regulating NTC checked by the service staff E...

Страница 18: ...530 1190 0 Jet connection short straight 80 mm 508 0427 0 Splash protection 501 0331 0 Rubber bush for splash protection 501 0331 1 Three holes gasket device side 501 0320 0 Three holes gasket handle side 530 1186 0 Hose double barrel for irrigation handle 530 1181 0 Jet connection for water irrigation handle 40 mm 502 0984 0 Hygiene filter for water irrigation 000 0918 0 Hose tips for jet connect...

Страница 19: ...or reducing bacterial count Irrigation mode water flow 0 450 ml min at 37 C Stimulation mode water flow 150 ml min at temperature preselected for the cold stimulation level resp warm stimulation level Heating off energy saving mode adjustable by timer from 1 up to 99 sec Accuracy 0 5 s digit 20 C 47 C Accuracy 0 8 C digit can only be guarantee with unchanged ambient conditions 7 19 5 C Drinking wa...

Страница 20: ...cle it must be disposed of professional According to the EC directives WEEE and RoHS the device may not be disposed of in domestic waste Please observe existing national laws and rules for disposal of old devices Disposal within the Federal Republic of Germany In the Federal Republic of Germany the law for electrical devices ElektroG rules the disposal of electrical devices In order to guarantee a...

Страница 21: ......

Страница 22: ...ds damage or loss is therefore transferred to the client as soon as the goods leave the factory or the client is in default of acceptance of the goods This also applies to cases where we confirm prepaid carriage Transport packaging and all other packaging according to the packaging regulations is not returnable Our client is responsible for disposing the packaging at its own cost Our deliveries ar...

Страница 23: ......

Страница 24: ......

Отзывы: