7.0 Trouble shooting
"E0"
not used
•
Check, whether the water supply delivers a pressure of at
least 2 bar (Did you open the water-tap?).
•
Filter clogged?
"E1"
no water (water pressure <0.5 bar)
•
Inform service staff.
"E2"
-5V is missing (supply voltage on the
control board)
•
Inform service staff.
"E3"
break of the safety NTC
"E4"
break of the safety NTC
•
Inform service staff.
"E5"
break of the regulating NTC
"E6"
not used
•
Check, whether temperature setting is too high.
If necessary, adjust desired temperature to a value of <48°C
by means of the coding switches.
•
Inform service staff.
"E7"
temperature too high (>48°C),
display only in the stimulation and
irrigation mode, but not in the
thermal process for reducing
bacterial count
Error in temperature display
•
Have temperature feeler of the regulating NTC checked by the
service staff.
"E8"
short-circuit of the regulating NTC
"E9"
not used
If, nevertheless, the errors cannot be removed, inform the ATMOS service staff. Do not start any attempts to repair the unit
yourself!
•
Inform service staff.