22
12.0
Technical specifi cations
Issue of Technical Specifi cations: 16.01.2019
Air fl ow rate of pump
22 ± 2 l/min
Max. vacuum
-76 kPa* ( -760 mbar; -570 mmHg) ±3kPa
Vacuum readout (for ATMOS
®
C 161 Battery)
-1...0 bar (± 25 mbar) (mmHg; kPa)*
Additional air regulation (for ATMOS
®
A 161 Bat-
tery)
Three-steps-adjustment
-25 kPa
*
, -55 kPa
*
,-75kPa
*.
* depends on daily atmospheric pressure and Ambient conditions
Additional air regulation (for ATMOS
®
C 161 Bat-
tery)
mechanical regulating valve
Collection canister
1 l DDS secretion canister, 1.5 l Receptal
®
canister system, 1 l Medi-Vac
®
canister
system or 1 l Serres
®
canister system
Suction hose
ø 6 mm, 1.30 m length
Voltage supply ((battery charging power supply))
100-240 VAC (+/-10%) 50/60 Hz
Charger
manufacturer: GlobTek Inc.
model: GTM91099-6015-3.0-T2
Low voltage supply
12 V DC ± 10%
Operation time during battery operation (without
mains supply)
Battery operation approx. 60 min.
Intermittent duty after 10 min (battery protection)
Operation time during mains supply 12 V DC (with
230 V battery charging power supply or 12 V batte-
ry charger leads)
Intermittent duty after 10 min (30 min. cooling down, depending on ambient tempe-
rature)
Emergency operation
In case of fully discharged battery, mains operation is possible
Power input (battery charging power supply)
max. 1.5 A
Power consumption
45 AV
Battery
8,4 V: 6 Ah; lithium-ionic at least 500 charging cycles, in case of partial discharge
accordingly more
Charging time
approx. 2 h 30 min
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
—
—
—
—
Heat emission
40 J/s
Noise level
56.0 dB (A) @ 1m (acc. to ISO 7779)
Ambient conditions
Transport/storage
Operation
-30...+50°C
5...90 % humidity,
non-condensing air pressure 700...1060 hPa
+10...+35°C
20...80 % humidity,
non-condensing air pressure 700...1060 hPa
Dimensions HxBxT
250 x 255 x 180 mm
Weight
4 kg
Period tests
Recommended: Testing every 24 months.
Protection class (EN 60601-1)
II
Application part
Type BF
Degree of protection
IP 21
Classifi cation acc. to Annex IX,
EC directive 93/42/EEC
IIa
CE marking
CE 0124
UMDNS-Code
10-219 (tracheal suction device)
GMDN-Code
37783
* 1 bar ≈ 750,06 mm Hg ≈ 1000 hPa / depends on daily atmospheric pressure