10
4.0 Cleaning
Do not use any abrasive cleaning agents. For the upholstery usual dry foam may be used.
Treatment with a commonly used care product for artifi cial leather is recommended once a week in order to keep the
upholstery soft and smooth.
!
4.2 Cleaning the unit's surfaces and upholstery
●
The surfaces of the ATMOS Chair 51 D
are resistant against all the recommended surface disinfectants stated in chapter 4.3
and 4.4. Nevertheless after any length of time discolourations could possibly develop.
●
Substances such as blood need to be removed immediately to prevent stains on the cover material of the upholstery.
●
The unit itself can be wiped off with a moist (not wet) cloth.
The described measures to clean and to disinfect do not
rule out regulations valid for the particular company!
●
In chapter 4.3 Recommended disinfectants“ all sur-
face and instrument disintectants are stated which are
suitable for disinfection.
Always observe the concentration specifi cations and
instructions by the respective manufacturer.
●
Do NOT use
- disinfectants containing concentrated organic
or anorganic acids or bases, since these may cause
corrosion damages.
- disinfectants containing chloramides, phenol
derivatives or anionic tensides, since these may
cause stress cracks in the plastics used.
4.1 General information on cleaning and disinfection
Before cleaning
Medical devices must always offer a maximum in safety and function.
4.3 Recommended surface disinfectant
● The surfaces of the ATMOS Chair 51 D can be cleaned/wiped with disinfectants containing the following active ingredients:
• QAV (quarternary ammonium compounds)
4.4 Recommended disinfectants for the upholstery
Disinfectant
Ingredients
(in 100 g) Manufacturer
ATMOS Green & Clean SK
Alkyldimethyl
(application solution)
benzylammonium
chloride
<
1
g
Metasys,
Rum
(Austria)
Dialkyldimehty
lammoniumchloride
<
1
g
Alkyldimethylethyl
benzylammoniumchloride
<
1
g