3.0 Setting up
Set up the device on a level, fi rm surface.
3.1
Operating elements
1
Shove blind for covering the operating elements
(ATMOS
®
C 161 / 261 Aspirator)
2
Vacuum adjustment
3
Vacuum gauge
(ATMOS
®
C 161 / 261 Aspirator)
4
Hose storage (fi g. 1, page 4)
5
Suction hose storage (accessory)
6
Switch On
I
/ Off
O
3.2 Connection
The main voltage specifi ed on the type plate must match
the power supply system.
Check mains cable for damages. Damaged cables must be
replaced immediately!
3.0 Setting up and starting
3.3 Starting up
●
The ATMOS
®
A / C 161 / 261 Aspirator is delivered
ready for use.
●
Lift the unit out of the cardboard. Check whether the
voltage values on the data plate correspond with the
inbuilding voltage.
●
Set up the device on a level, fi rm surface.
●
Prior to fi rst operation, pay attention to the safety
information in chapter 2.0.
●
The unit must be allowed to stand for up to six hours
at room temperature prior to starting up for the fi rst
time following transport at temperatures below
freezing. The unit may
not
be operated if it has
not
acclimatised as this might damage its diaphragms.
●
ATMOS
®
A / C 161 / 261 Aspirator / DDS: Always hold
ready at least one DDS bacterial filter, as the device
may not be operated without one!