18
10.0 Trouble-shooting
Prior to dispatch, the ATMOS
®
A / C 161 / 261 Aspirator was subjected to an extensive functional test. If, nevertheless, a failure
should appear, you may possibly clear it yourself if you follow these notes
:
Problem
Possible causes
Remedy
● Unit does not start
– Discharged battery
– Connect the battery charging
power pack to the device. The
battery should be recharged for
1 – 2 hours prior to operation
with battery.
– Loose power plug of the charging device – Check all plug and socket
connections. Pay attention to the
control lamp, it must be
illuminated when all connections
are o.k.
●
Insuffi cient performance
– Discharged battery
– Recharge the battery.
– Filter is blocked
– Exchange the fi lter.
● 1. Low or no vacuum
is indicated
– 1.1 DDS bacterial fi lter is missing
– Insert DDS bacterial fi lter
– 1.2 Leakages within the hose system
or in collection canister lid
– Check collection canister lid and
hose system on tight fi tting.
– Connect the fi lter once again to
the connection nozzle.
– Check the suction lid on tight
and correct fi tting.
● 2. High vacuum is
indicated
– 1.3 Secretion or blood has been
sucked in and valve plates of
the pump are contaminated
– 2.1 DDS bacterial fi lter is blocked.
– Unit has to be returned for repair.
– Exchange DDS bacterial fi lter
– 2.2 Float of overfl ow safety closes
the collection canister inlet
– Check collection canister inlet; if
necessary empty secretion
canister, clean the oversuction
protection and check fl oat ball
for fl exibility.