Operating Manual Xenotest
®
440
- 54 -
Inhalt/Content/Sommaire
10.3 Maintenance
WARNING – electrical shock!
Touching live parts can lead to life threat-
ening electrical shock. Disconnect the in-
strument from the power supply before be-
ginning maintenance work!
• Turn off the instrument at the main ON/OFF
switch
• Pull the electrical plug out of the socket
and secure against reconnecting
• Check whether the instrument is voltage
free
Cleaning the air filters:
The air filters for lamp cooling and test chamber cool-
ing must be cleaned every six months. Place the filter
in warm water with a mild detergent and rinse well. In
case of excessive dirt, tap the filter mat clean, vacuum
it or blow it out using soft compressed air, then clean it
in water.
Changing the air filters:
The two filters must be changed with new filters after
cleaning the existing filters twice.
Removing and installing the lamp cooling air
filter
Fig. 43:
1. Open the right hand instrument door
1
.
2. Pull the filter
3
to the right out of the holder
2
.
3. Push the filter into the holder.
4. Close the instrument door again.
Removing and installing the test chamber cool-
ing air filter
Fig. 44
1. Pull the filter
1
up out of the holder
2
.
2. Push the filter into the holder.
3
2
1
1
2
Fig. 43
Fig. 44
10
Maintenance