- 41 -
Operating Manual Xenotest
®
440
Inhalt/Content/Sommaire
6.7
Filling the water tank
Manual filling of the water tank
Fig. 33:
1. Pull open the bottom door
2
on the front of the in-
strument, which is held by a magnet.
2. Lift the lid
1
of the water tank
3
. Fill the water tank
with approx. 40 to 60 liters of water (see specification
of the treated water sect. 3.4).
3. Close the lid of the water tank and then the instru
-
ment door.
When connecting to an external water circuit:
1. Open the water supply.
2. Check the fill level in the water tank.
6.8
Connecting the exhaust air system
NOTE – exhaust air lamp cooling
It is highly recommended that the lamp cooling air be
connected to an exhaust air system. The instrument ex-
haust air system can overheat if the maximum length of
the exhaust air channel and the maximum permissible
number of two 90° bends is not observed. This may re
-
sult in the instrument exhaust system overheating. In
turn, this will cause the instrument safety equipment to
continuously interrupt the power supply to the instru-
ment. Please observe the specifications for the instru
-
ment exhaust air system in sect. 3.3.
CAUTION – harmful UV radiation!
If the instrument is not connected to an on-
site exhaust system, hazardous levels of
UV radiation may be emitted. UV radiation
can cause irritation to the retina and skin.
Only operate the instrument with a func-
tional exhaust air system installed.
Connecting exhaust air system Xenotest
®
440
Fig. 34:
1. Place the adapter
1
of the customer’s exhaust air
system on the connection nozzle
2
on Xenotest 440.
The connection nozzle has a diameter of 100 mm.
2. Check whether the exhaust air system has an open
output outdoors or the on-site technical ventilation
system is ready for operation.
3
2
1
Fig. 33
2
1
Fig. 34
6
Start up