9

6.   

Start Button:

 Starts treatment cycle time. 

Once selected:

a) Remaining Time changes to Countdown Time and begins counting down from 1:29 minutes to exit area

b) The words 

“LEAVE AREA”

 appear on the lower status bar. 

c) Alarm on unit begins to sound

d) The word 

“PREP”

 flashes on the Keypad during the Countdown Time, when Treatment Time begins the word 

“RUN”

 

flashes and when the treatment time is completed the word 

“DONE”

 will appear.

7.   

Return Button:

 Brings you back to previous screen

8.   

Pause / Continue Button:

 Used to pause the treatment time, or to continue the treatment time. This button toggles from the 

Pause Icon

, which is displayed as two vertical bars to the 

Continue Icon

 which is a looping arrow. If pause is selected below 

30 seconds during the Countdown, the Countdown time will reset to 30 seconds. If pause is selected during the cycle time the 

device will provide an additional 1:30 seconds of countdown time to leave the area. 

9.   

Battery Condition:

 Displays the charge level of the battery in the hand-held remote. Press on battery symbol to see 

 

percent of change.

10.   

Settings Button:

 Contains choices for:

a) Settings: Reset Operating Time or Clear Session Data

b) Set Date/Time: requires a PIN to reset (See 

“Setting Date and Time”

 on 

Page 12

)

c) Sessions: Log of saved sessions

11.   

Remaining Time

:

 Displays remaining time of treatment and toggles to (Countdown time shown on Treatment in Hours, 

Minutes, Seconds) when selected. 

12.   

Total Operation Time

:

 Displays accumulative hours of all treatment times.

13.   

Stop or End Button:

 Stop treatment time.

14.   

Status Bar:

 Outputs all the conditions of the device

a) Ready

b) Intruder Alert

c) Session in Progress

d) Session Paused

e) Session Resumed

f) Session Stopped

g) Leave Area

h) Connected to 

Sanidyne

®

15.   

Settings Button:

 Same as (10) above.

16.   

Home Button

:

 Exit Setting 

l

amP

 i

nstallation

 

  

Always disconnect power to the fixture before performing any service or maintenance

.

Germicidal lamps are easily damaged. Exercise care when handling lamps.

Always Disconnect Power to the 

Sanidyne

®

 

Prime Remote 

before 

installing or removing lamps.

Step 1

 - Insert lamp end without pins 

through 

BOTTOM

 of lamp guard while 

holding each lamp end with one hand.

Step 2

 - Carefully align pins on lamp 

base to socket holes on lampholder 

as shown. 

(Lamp Installation 

Continued on Page 10.)

Содержание Sanidyne 40-0634A

Страница 1: ...duct Information Available at Ultraviolet com BuyUltraviolet com Read and Follow All Safety Instructions Save These Instructions Installation Operation Maintenance 8 High Output UV Lamps with ArmorLit...

Страница 2: ...NTENANCE LOG 17 19 USER ASSISTANCE 20 Patent Notice 20 WARRANTY PRODUCT REGISTRATION 20 These instructions generally describe the installation operation and maintenance of the Sanidyne Prime Remote Po...

Страница 3: ...3 Read and follow all notices and warnings on the sanitizer 14 SAVE THESE INSTRUCTIONS SAFETY WARNINGS All personnel should be alerted to the potential hazards indicated by the product safety labeling...

Страница 4: ...equipment for use in the user s system or process No warranty or representation is made by the manufacturer with respect to suitability or performance of this equipment or to the results that may be...

Страница 5: ...ovide direct germicidal ultraviolet exposure for air and surface disinfection in unoccupied areas to disinfect and prevent areas and surfaces from microorganisms Ultraviolet disinfection dosage is a f...

Страница 6: ...de until ready for installation Carefully inspect all packing materials to prevent the loss of accessories spare parts or instructions Do not discard packing materials until unit is assembled checked...

Страница 7: ...is completed 2 Hour Buttons Up and down buttons are used to select treatment hours between 0 and 24 3 Minute Buttons Up and down buttons are used to select treatment minutes between 0 and 59 4 Start...

Страница 8: ...s three icons at bottom which are Sound Off Vibrate Sound On 2 Indicator Light Glows RED when the device begins to turn on and during device charging Turns GREEN when fully charged 3 Treatment Time Se...

Страница 9: ...to see percent of change 10 Settings Button Contains choices for a Settings Reset Operating Time or Clear Session Data b Set Date Time requires a PIN to reset See Setting Date and Time on Page 12 c S...

Страница 10: ...ns supplied with the Sanidyne Prime Remote on all access doors to the area where treatment operation will take place It is the users responsibility to post signs or notices for all personnel and perso...

Страница 11: ...econds remain on the countdown time the Sanidyne Prime Remote will automatically reset giving 30 seconds more to leave the Operation Area Occupancy Sensors do not protect the entire Operation Area of...

Страница 12: ...someone near the Sanidyne Prime Remote the unit will immediately turn off and send an alarm to the Hand held Remote Controller and the Sanidyne Prime Remote If this event occurs the alarm can be turne...

Страница 13: ...e Remote on each side of large fixed objects like a bed or table etc 2 Measure the longest distance from the Sanidyne Prime Remote to the farthest wall Use this length to compare to the Distance From...

Страница 14: ...S BETWEEN CUFFS AND GLOVES AND AN ULTRAVIOLET BLOCKING FACE SHIELD MUST BE WORN TO COVER AND PROTECT EYES AND EXPOSED SKIN NOTE STER L RAY lamps used in the Sanidyne Prime Remote have ArmorLiteTM Safe...

Страница 15: ...fied and factory authorized personnel Table 1 Troubleshooting Problem Possible Cause Corrective Action Sanidyne Prime Remote not operating Power disconnected Check connections Fuse not operating Remov...

Страница 16: ...ity Aging Measuring UV C Germicidal Fixture Leakage 2 Promate Safety Glasses Safety eyewear MUST be used as general purpose safety protection and for additional shielding from ultraviolet rays 3 Addit...

Страница 17: ...Part No Part No 1 STER L RAY Germicidal Lamp 05 0892 05 0892 2 SureliteTM Ballast 10 0138 10 0138 3 Power Cord 35 1754 35 1767 4 PromateTM Face Shield Adjustable Headgear 00 0126 00 0126 5 PromateTM W...

Страница 18: ...18 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Страница 19: ...19 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Страница 20: ...g warranties of merchantability and fitness for a particular purpose whether or not purposes or specifications are described herein We further disclaim any responsibility whatsoever to the customer or...

Отзывы: