background image

11

s

aniDyne

®

 P

rime

 r

emote

 o

Peration

The 

Sanidyne

®

 

Prime Remote

 has two methods of controlling the unit; the Keypad Controller (Shown in 

Figure 2

) and the 

Hand-held Remote Controller (Shown in 

Figure 3

). These two controls mirror each other; treatment times entered into Keypad 

Controller will appear on the Hand-held Remote Controller and vice versa.

u

sing

 

tHe

 K

eyPaD

 C

ontroller

 

to

 s

et

 t

reatment

 t

imes

1. 

Setting Treatment Time on the Keypad is done using the 

HOURS

 and 

MINUTES

 buttons which are located on the top of 

the unit. (See 

Figure 1

 for location) Selecting the up or down facing arrows located on the keypad in line with the 

HOURS

 

or 

MINUTES

 button will either increase or decrease the hours or minutes. The hours can be set between 0 and 24, and 

the minutes can be set from 1 to 59. (Example of one hour and five minutes would appear as 01:05 on the Treatment Time 

Display). Since the Keypad Controller will no longer be viewable once the treatment time begins, it is recommended that the 

Hand-held Remote Controller be fully charged and powered before this time and be in possession of the operator at all times.

2. 

Determining the proper treatment times for your specific location can be done by using the 

“Recommended Treatment Time 

for Room Size”

 chart (

Figure 4

 on 

Page 13

). 

3. 

Once the choice of time has been displayed on the Keypad controller, select the 

START/STOP

 button to begin and leave the 

Operation Area immediately. Make sure that the Hand-held Remote Controller is kept in possession of the operator during the 

Treatment Time.

Once you select the START/STOP

 button an audible exit alarm will begin to sound and the 

Sanidyne

®

 

Prime 

Remote

 will begin counting down from 1:30 seconds to allow all inhabitants to leave the area before the UV 

treatment begins. Additionally the word “

PREP

” will begin flashing on the Keypad. The occupancy Sensors located 

on the top of the unit (Shown on Figure 1) provide an additional safety area around the unit will begin to operate. As 

an additional safety feature, if the sensors detect movement close to the 

Sanidyne

®

 

Prime Remote

 when less than 

30 seconds remain on the countdown time, the 

Sanidyne

®

 

Prime Remote

 will automatically reset giving 30 seconds 

more to leave the Operation Area.

Occupancy Sensors do not protect the entire Operation Area of the 

Sanidyne

®

 

Prime Remote

. Under no 

circumstances should this sanitizer be permitted to operate with humans, plants, or animals present in the Operation Area

.

4. 

Outside the Operation Area the Treatment Time can be viewed on the Hand-held Remote Controller. Once the treatment time 

has expired the unit will turn off the 

STER-L-RAY

®

 Germicidal Ultraviolet Lamps and it is now safe to return to the area. 

The word 

“DONE” 

will begin flashing on the Keypad Controller on the top of the unit. 

u

sing

 

tHe

 H

anD

-

HelD

 r

emote

 C

ontroller

 

to

 s

et

 t

reatment

 t

imes

The Hand-held Remote Controller provides further detailed information on the 

Sanidyne

®

 

Prime Remote

. Therefore it is 

recommended to review the Hand-held Remote Controller Overview section of this manual to become familiar with the display 

which is shown in 

Figure 3

The Hand-held Remote Controller can operate the 

Sanidyne

®

 

Prime Remote

 remotely, thus it is 

imperative that the operator make a visual inspection of the Operation Area to ensure it is clear of all humans, plants, or 

animals before proceeding.

1. 

Setting Treatment Time on the Hand-held Remote Controller is done by holding a finger on the display face, directly on the 

words Treatment Time until the screen changes to a scroll bar. Next scroll the numbers up or down, characters on the left are 

hours, characters on the right are minutes. Positioning the desired treatment time in the center row. (Example of one hour and 

five minutes would appear as 01:05 on the Treatment Time Display) When times are positioned correctly, select 

DONE

 on the 

bottom of the scroll bar to input your Treatment Time choice.

2. 

Determining the proper treatment times for your specific location can be done by using the 

“Recommended Treatment Time 

for Room Size”

 chart (

Figure 4

 on 

Page 13

). 

3. 

Once the choice of time has been displayed on the Hand-held Remote Controller, select the start button to begin and leave the 

Operation Area immediately. Make sure that the Hand-held Remote Controller is kept in possession of the operator during the 

Treatment Time.  

(Remote Controller Instructions Continued on Page 12)

Содержание Sanidyne 40-0634A

Страница 1: ...duct Information Available at Ultraviolet com BuyUltraviolet com Read and Follow All Safety Instructions Save These Instructions Installation Operation Maintenance 8 High Output UV Lamps with ArmorLit...

Страница 2: ...NTENANCE LOG 17 19 USER ASSISTANCE 20 Patent Notice 20 WARRANTY PRODUCT REGISTRATION 20 These instructions generally describe the installation operation and maintenance of the Sanidyne Prime Remote Po...

Страница 3: ...3 Read and follow all notices and warnings on the sanitizer 14 SAVE THESE INSTRUCTIONS SAFETY WARNINGS All personnel should be alerted to the potential hazards indicated by the product safety labeling...

Страница 4: ...equipment for use in the user s system or process No warranty or representation is made by the manufacturer with respect to suitability or performance of this equipment or to the results that may be...

Страница 5: ...ovide direct germicidal ultraviolet exposure for air and surface disinfection in unoccupied areas to disinfect and prevent areas and surfaces from microorganisms Ultraviolet disinfection dosage is a f...

Страница 6: ...de until ready for installation Carefully inspect all packing materials to prevent the loss of accessories spare parts or instructions Do not discard packing materials until unit is assembled checked...

Страница 7: ...is completed 2 Hour Buttons Up and down buttons are used to select treatment hours between 0 and 24 3 Minute Buttons Up and down buttons are used to select treatment minutes between 0 and 59 4 Start...

Страница 8: ...s three icons at bottom which are Sound Off Vibrate Sound On 2 Indicator Light Glows RED when the device begins to turn on and during device charging Turns GREEN when fully charged 3 Treatment Time Se...

Страница 9: ...to see percent of change 10 Settings Button Contains choices for a Settings Reset Operating Time or Clear Session Data b Set Date Time requires a PIN to reset See Setting Date and Time on Page 12 c S...

Страница 10: ...ns supplied with the Sanidyne Prime Remote on all access doors to the area where treatment operation will take place It is the users responsibility to post signs or notices for all personnel and perso...

Страница 11: ...econds remain on the countdown time the Sanidyne Prime Remote will automatically reset giving 30 seconds more to leave the Operation Area Occupancy Sensors do not protect the entire Operation Area of...

Страница 12: ...someone near the Sanidyne Prime Remote the unit will immediately turn off and send an alarm to the Hand held Remote Controller and the Sanidyne Prime Remote If this event occurs the alarm can be turne...

Страница 13: ...e Remote on each side of large fixed objects like a bed or table etc 2 Measure the longest distance from the Sanidyne Prime Remote to the farthest wall Use this length to compare to the Distance From...

Страница 14: ...S BETWEEN CUFFS AND GLOVES AND AN ULTRAVIOLET BLOCKING FACE SHIELD MUST BE WORN TO COVER AND PROTECT EYES AND EXPOSED SKIN NOTE STER L RAY lamps used in the Sanidyne Prime Remote have ArmorLiteTM Safe...

Страница 15: ...fied and factory authorized personnel Table 1 Troubleshooting Problem Possible Cause Corrective Action Sanidyne Prime Remote not operating Power disconnected Check connections Fuse not operating Remov...

Страница 16: ...ity Aging Measuring UV C Germicidal Fixture Leakage 2 Promate Safety Glasses Safety eyewear MUST be used as general purpose safety protection and for additional shielding from ultraviolet rays 3 Addit...

Страница 17: ...Part No Part No 1 STER L RAY Germicidal Lamp 05 0892 05 0892 2 SureliteTM Ballast 10 0138 10 0138 3 Power Cord 35 1754 35 1767 4 PromateTM Face Shield Adjustable Headgear 00 0126 00 0126 5 PromateTM W...

Страница 18: ...18 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Страница 19: ...19 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Страница 20: ...g warranties of merchantability and fitness for a particular purpose whether or not purposes or specifications are described herein We further disclaim any responsibility whatsoever to the customer or...