background image

 

 

47 

O&M Manual 
Rev-H (6/17) 

Part 9 – System Maintenance 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

9.1 

General 

 

 

The  Q46/84  Hydrogen  Peroxide  System  will  generally  provide  unattended  operation  over  long 
periods  of  time.    With  proper  care,  the  system  should  continue  to  provide  measurements 
indefinitely.    For  reliable  operation,  maintenance  on  the  system  must  be  done  on  a  regular 
schedule.    Keep  in  mind  that  preventive  maintenance  on  a  regular  schedule  is  much  less 
troublesome than emergency maintenance that always seems to come at the wrong time.   

 

9.2 

Analyzer Maintenance 

 

 

No unusual maintenance of the analyzer is required if installed according to the guidelines of this 
operating  manual.    If  the  enclosure  door  is  frequently  opened  and  closed,  it  would  be  wise  to 
periodically inspect the enclosure sealing gasket for breaks or tears. 

 

9.3 

Peroxide Sensor Maintenance 

 

Sensor maintenance is required for accurate measurements.  The primary requirement is simply 
to keep the sensor membrane clean.  The membrane is a micro-porous polymer that is resistant 
to anything that will be encountered in water streams.  However, deposits can form on the surface 
or  in  the  pores  of  the  membrane,  and  these  deposits  will  reduce  the  sensitivity.    Certain 
constituents in water, mainly iron and manganese, will form precipitates.  These precipitates can 
sometimes form a coating on the membrane.   
 
Because  membranes  are  micro-porous,  they  can  be  relatively  difficult  to  clean  effectively.  
Immersing  the  tip  of  the  sensor  in  1N  nitric  acid  solution  will  sometimes  remove  deposits  that 
cause  low  sensitivity,  but  this  is  not  always  the  case.    The  recommended  practice  is  to  simply 
replace  the  membrane  when  it  becomes  fouled.    To  change  a  membrane,  follow  the  Sensor 
Assembly  procedure  on  page  14  of  this  manual.    Do  not  reuse  the  electrolyte  from  the  sensor 
when  changing  a  membrane.    Always  refill  with  fresh  electrolyte.    The  electrolyte  is  stable  and 
does not have a limited shelf life. 

 

Refer again to the explanation of the sensor slope number after an accepted span calibration on 
the  lower  MEASURE  screen.    In  normal  operation,  the  slope  of  the  sensor  output  will  decrease 
over  time  as  the  membrane  becomes  fouled.    This  reduction  indicates  that  the  sensor  is  losing 
sensitivity  to H

2

O

2

.  It  is good practice to replace the  membrane if the slope number falls to 30-

40%.  The value will not go below 20%. 

 

Even  if  no  buildup  is  apparent  on  the  membrane,  it  should  be  changed  on  a  regular  schedule.  
The  recommended  membrane  change  interval  is  every  3  months.    For  high  purity  water 
applications, this can probably be extended if desired, but a more frequent changing interval is a 
small price to pay for avoiding membrane failure at the wrong time. 
 
While the sensor  is disassembled for membrane changing,  examine the condition of the o-rings 
on both ends of the electrolyte chamber.  If the o-rings show any signs of damage, replace them 
with new ones from the spare parts kit.  It is good practice to change these o-rings once a year, 
regardless of their condition.  

 

Содержание Q46/84

Страница 1: ...ology Inc ATI UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Phone 800 959 0299 Saddleworth OL3 5DE 610 917 0991 Phone 44 0 1457 873 318 Fax...

Страница 2: ...2 Software 26 6 21 Software Navigation 27 6 22 Measure Menu MEASURE 29 6 23 Calibration Menu CAL 30 6 24 Configuration Menu CONFIG 30 6 25 Control Menu CONTROL 33 6 26 Diagnostics Menu DIAG 37 PART 7...

Страница 3: ...WCELL DETAILS 14 FIGURE 9 LINE POWER CONNECTION 16 FIGURE 10 RELAY CONTACTS 17 FIGURE 11 OPTIONAL RELAY BOARD WIRING 18 FIGURE 12 OPTIONAL ANALOG OUTPUT WIRING 18 FIGURE 13 SENSOR CABLE PREPARATION 19...

Страница 4: ...d in the monitor is a polarographic membraned sensor which measures hydrogen peroxide directly Water simply flows past the sensor and directly to drain with the flow rate and pressure across the senso...

Страница 5: ...ation For connection of the sensor to the electronics a 25 cable is supplied An additional 100 feet of interconnect cable may be added using 07 0100 junction box All required spare parts are also prov...

Страница 6: ...Selectable PID controller on main analog output PID controller can operate with instrument configured as loop power transmitter or as one of the two outputs on the AC powered instrument PID includes...

Страница 7: ...with sensor flowcell Ambient Temperature Analyzer Service 20 to 60 C 4 to 140 F Sensor Service 5 to 55 C 23 to 131 F Storage 5 to 70 C 22 to 158 F Ambient Humidity 0 to 95 non condensing EMI RFI Infl...

Страница 8: ...ptional Relays Three additional non isolated low power relays available SPST 1 Amp 30 VDC Analog Outputs Two 4 20 mA outputs Output one programmable for PPM peroxide or PID Output 2 programmable for P...

Страница 9: ...instrument in close proximity to the point of sensor installation this will allow easy access during calibration The sensor to instrument distance should not exceed 100 feet To maximize signal to nois...

Страница 10: ...instrument is then attached to the wall using the four outer mounting holes in the bracket These holes are slotted to accommodate two sizes of u bolt that may be used to pipe mount the unit Slots wil...

Страница 11: ...ATI Q46 84 Hydrogen Peroxide Monitor Part 2 Analyzer Mounting 11 O M Manual Rev H 6 17 Figure 4 Wall Mounting Diagram Figure 5 Pipe Mounting Diagram...

Страница 12: ...tems The panel mounting bracket kit must be ordered separately part 05 0068 This kit contains a metal retainer bracket that attaches to the rear of the enclosure 4 screws for attachment of this bracke...

Страница 13: ...w concentrations below 0 5 PPM this method should always be used Mechanical installation of the flow cell requires that it be mounted to a wall or other convenient flat surface Alternatively the mount...

Страница 14: ...let flow is well controlled can use a simpler sealed flowcell Using this flowcell requires that flow be controlled externally to about 400 cc min Variable flow rate or variable pressure will cause uns...

Страница 15: ...relevant local codes and instructions contained in this operating manual Observe the analyzer s technical specifications and input ratings Proper electrical disconnection means must be provided prior...

Страница 16: ...the microprocessor assembly located in the front section of the enclosure This cable may be unplugged from the front section of the monitor if service is needed but should normally be left in place d...

Страница 17: ...e proper power to the contacts For applications that require the same switched operating voltage as the Q46 115 or 230 V power may be jumpered from the power input terminals at TB7 Relay wiring is con...

Страница 18: ...o connect to the optional 3 relay card Figures 11 OR the optional third analog output Out 3 Figure 12 The Q46 can be configured for only one of these optional features and the hardware for either opti...

Страница 19: ...removed easily from the flowcell Cord grips used for sealing the cable should be snugly tightened after electrical connections have been made to prevent moisture incursion When stripping cables leave...

Страница 20: ...o available to provide a break point for long sensor cable runs Route signal cable away from AC power lines adjustable frequency drives motors or other noisy electrical signal lines Do not run sensor...

Страница 21: ...ax 100 feet a junction box must be used The junction box is shown in Figure 15 and is supplied with a conduit hub on one end and a sensor cable gland on the other end NOTE If sensor is experiencing Lo...

Страница 22: ...2O2 Sensor Assembly Follow the procedure below to prepare the sensor for operation CAUTION Use extreme care when filling the H2O2 sensor with electrolyte The electrolyte is very acidic and can cause e...

Страница 23: ...lowly screw the chamber onto the sensor body A small amount of electrolyte will run out of the hole from which the vent screw was removed Place a paper towel around the sensor to absorb the electrolyt...

Страница 24: ...from this user interface no internal jumpers pots etc Figure 17 User Interface MENU ICONS UNITS 12 CHARACTER SECONDARY DISPLAY MEMBRANE KEYPAD MENU ESC ENTER A B DIAG FAIL HOLD CAL CONF MENU ICONS UN...

Страница 25: ...tem for change and to store any change 6 12 Display The large custom display provides clear information for general measurement use and user configuration There are three main areas of the display the...

Страница 26: ...the output is locked to the last input value measured when the HOLD function was entered HOLD values are retained even if the unit power is cycled FAIL The FAIL icon indicates that the system diagnost...

Страница 27: ...ompleted pressing ENTER again stores the variable and stops the flashing Pressing ESC aborts the change and also exits user entry mode The starting default screen is always the Measure Menu The UP arr...

Страница 28: ...AIL Relay B CON Set Default Setpnt A Delay A Phase A Hyst A Relay A AL Relay A FAIL Relay A CON Setpnt A HI Delay A HI Setpnt A LO Hyst A HI Delay A LO Phase A Hyst A LO Phase B Slope CONTROL Fail Val...

Страница 29: ...er can be placed in manual while viewing this screen by pressing and holding the ENTER key for 5 seconds until a small flashing m appears on the screen At that point the controller output can be adjus...

Страница 30: ...ensitivity is to specification in any user selected range 6 24 Configuration Menu CONFIG The Configuration Menu contains all of the general user settings Entry Lock This function allows the user to lo...

Страница 31: ...alue will flash Use the UP arrow key to modify the desired display value Press ENTER to store the new value Com Mode Sets digital communication mode of analyzer Optional digital communication card mus...

Страница 32: ...B and C can be used in two ways as a setpoint control or as an Relay C Mode alarm The two settings for Relay B Mode are CON and FAIL The CON setting enables normal setpoint operation for Relay B C Re...

Страница 33: ...pting to maintain by Iout1 PID adjusting output value It is the nature of the PID controller that it never actually gets to the exact value and stops The controller is continually making smaller and s...

Страница 34: ...tions set the optional third 4 mA and 20 mA Set 20 mA 3 current loop output points for the analyzer The output may be set to temp H2O2 track temperature default or peroxide The values stored for the 4...

Страница 35: ...LO the relay energizes and the LCD indicator illuminates when the H2O2 level drops below the setpoint The failsafe setting does have an impact on this logic The description here assumes the failsafe s...

Страница 36: ...st automatically If Relay C Mode is set to CON see Relay C Mode then Relay C will function identically to Relay A Relay C settings appear in the CONFIG menu list automatically If Relay D E or F Mode i...

Страница 37: ...tion on Press the ESC key to go to the DIAG Menu and scroll to Sim Output using the UP arrow key Press ENTER Follow the instructions under Sim Output see following page CAUTION There is no time out on...

Страница 38: ...ontrol of the SIM screen until the ESC key is pressed Note If the HOLD function is engaged before the Sim Output function is engaged the simulated output will remain the same even when the ESC key is...

Страница 39: ...F under this normal condition Therefore when the trip condition occurs the closed N O contacts will be opened relay de energized and the LCD icon will illuminate In addition a power fail would also ca...

Страница 40: ...d need not be distilled but it must not contain hydrogen peroxide If the sensor is being used for the first time allow the sensor to sit undisturbed for at least 4 hours If the sensor has been in use...

Страница 41: ...the sensor into the flow cell until the membrane cap bottoms out on the acrylic flow cell Do not over tighten 2 Turn on the inlet water flow to the flow cell and adjust the inlet flow rate so that wa...

Страница 42: ...ibrium with the solution in order to provide the highest accuracy When moving the sensor between widely different temperature conditions it may be necessary to allow the sensor to stabilize as much as...

Страница 43: ...chemicals etc that enable the controller to add or subtract chemical rapidly enough This is not only specific to pumps and valves but also to line sizes delays in the system etc This section is includ...

Страница 44: ...ber between 3 5 to start 1 2 more than P Like proportional gain increasing the integral term will cause the controller action to be quicker D Derivative gain The addition of derivative control can be...

Страница 45: ...and hold the ENTER key for about 5 seconds You will see a small m show up between the value and the mA value This indicates you are now in manual mode Once in manual you may increase the PID output by...

Страница 46: ...al at the extreme ranges of the measuring scale In addition if a chemical process is not symmetrical that means it responds differentially to the addition and subtraction of chemical It is important i...

Страница 47: ...ifficult to clean effectively Immersing the tip of the sensor in 1N nitric acid solution will sometimes remove deposits that cause low sensitivity but this is not always the case The recommended pract...

Страница 48: ...f the sensor to be submerged Allow the electrodes to soak in distilled water for 30 minutes 4 Put a new membrane and fresh electrolyte in the electrolyte chamber and reassemble the sensor Connect to t...

Страница 49: ...4 20 mA outputs do not exceed the limits in the Instrument Specifications During troubleshooting it is many times helpful to disconnect all these outputs and place wire shorts across the terminals in...

Страница 50: ...monitor to verify that the millivolt value is actually 400 mV 4 Check TC drive circuit Place a wire short between the GREEN and BLACK terminals With a DVM measure the voltage between terminals BLACK a...

Страница 51: ...above TC Error TC may be open or shorted Check sensor wiring and perform RTD test as described in sensor manual Check j box connections Figure 22 Display Messages MESSAGE DESCRIPTION POSSIBLE CORRECTI...

Страница 52: ...e red and white wires The red wire is part of the RTD circuit and the white wire is part of the measuring cell There should be no connection Reading resistance you should find an open circuit value of...

Страница 53: ...C with 3 rd 4 20mA output 03 0409 Q46 P S Assy 100 240 VAC with 3 relay exp Board 03 0410 Q46 P S Assy 12 24 VDC 03 0411 Q46 P S Assy 12 24 VDC with 3 rd 4 20mA output 03 0412 Q46 P S Assy 12 24 VDC w...

Страница 54: ...of extended exposure to excessive gas concentrations Should inspection indicate that sensors have been expended rather than failed prematurely the warranty shall not apply The Manufacturer assumes no...

Страница 55: ...DUCTS NH3 Ammonia CO Carbon Monoxide H2 Hydrogen NO Nitric Oxide O2 Oxygen CO Cl2 Phosgene Br2 Bromine Cl2 Chlorine ClO2 Chlorine Dioxide F2 Fluorine I2 Iodine HX Acid Gases C2H4O Ethylene Oxide C2H6O...

Отзывы: