background image

 

 

43 

O&M Manual 
Rev-H (6/17) 

Part 8 – PID Controller Details 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8.1 

PID Description 

 

 

PID  control,  like  many  other  control  schemes,  are  used  in  chemical  control  to  improve  the 
efficiency  of  chemical  addition  or  control.    By  properly  tuning  the  control  loop  that  controls 
chemical  addition,  only  the  amount  of  chemical  that  is  truly  required  is  added  to  the  system, 
saving money.  The savings can be substantial when compared to a system which may be simply 
adding chemical at a constant rate to maintain some minimal addition under even the worst case 
conditions.    The  PID  output  controller  is  highly  advantageous  over  simple  control  schemes  that 
just  utilize  direct  (proportional  only)  4-20  mA  output  connections  for  control,    since  the  PID 
controller can automatically adjust the “rate” of recovery based on the error between the setpoint 
and the measured value – which can be a substantial efficiency improvement..   
 
 

 

The  PID  controller  is  basically  designed  to  provide  a  “servo”  action  on  the  4-20 

mA  output  to  control  a  process.    If  the  user  requires  that  a  measured  process  stay  as  close  as 
possible  to    a  specific  setpoint  value,  the  controller  output  will  change  from  0%  to  100%  in  an 
effort to keep the process at the setpoint.  To effect this control, the controller must be used with 
properly  selected  control  elements  (valves,  proper  chemicals,  etc.)  that  enable  the  controller  to 
add or subtract chemical rapidly enough.  This is not only specific to pumps and valves, but also 
to line sizes, delays in the system, etc.  
          

  

 

 

This  section  is  included  to  give  a  brief  description  of  tuning  details  for  the  PID 

controller, and is not intended to be an exhaustive analysis of the complexities of PID loop tuning.  
Numerous  sources  are  available  for  specialized  methods  of  tuning  that  are  appropriate  for  a 
specific application.   

 

 

8.2 

PID Algorithm 

 

 

As  most  users  of  PID  controllers  realize,  the  terminology  for  the  actual  algorithm  terms 

and even the algorithms themselves can vary between different manufacturers. This is important 
to  recognize  as  early  as  possible,  since  just  plugging  in  similar  values  from  one  controller  into 
another  can  result  in  dramatically  different  results.    There  are  various    basic  forms  of  PID 
algorithms  that  are  commonly  seen,  and  the  implementation  here  is  the  most  common  version; 
The ISA algorithm (commonly referred to as the “ideal” algorithm.) 





+

+

=

dt

t

de

D

t

d

t

e

I

t

e

P

output

)

(

)

(

)

(

1

)

(

 

 

 

Where: 

 

  

output = 

controller output 

 

 

P = 

proportional gain 

 

 

I = 

integral gain 

 

 

D = 

derivative gain 

 

 

t = 

time 

 

 

e(t) = 

controller error (e=measured variable – setpoint) 

 

Figure 21 - ISA (Ideal) Equation 

The  most  notable  feature  of  the  algorithm  is  the  fact  the  proportional  gain  term  affects  all 
components  directly  (unlike  some  other  algorithms  -  like  the  “series”  form.)    If  a  pre-existing 
controller utilizes the same form of the algorithm shown above, it is likely similar settings can for 

Содержание Q46/84

Страница 1: ...ology Inc ATI UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Phone 800 959 0299 Saddleworth OL3 5DE 610 917 0991 Phone 44 0 1457 873 318 Fax...

Страница 2: ...2 Software 26 6 21 Software Navigation 27 6 22 Measure Menu MEASURE 29 6 23 Calibration Menu CAL 30 6 24 Configuration Menu CONFIG 30 6 25 Control Menu CONTROL 33 6 26 Diagnostics Menu DIAG 37 PART 7...

Страница 3: ...WCELL DETAILS 14 FIGURE 9 LINE POWER CONNECTION 16 FIGURE 10 RELAY CONTACTS 17 FIGURE 11 OPTIONAL RELAY BOARD WIRING 18 FIGURE 12 OPTIONAL ANALOG OUTPUT WIRING 18 FIGURE 13 SENSOR CABLE PREPARATION 19...

Страница 4: ...d in the monitor is a polarographic membraned sensor which measures hydrogen peroxide directly Water simply flows past the sensor and directly to drain with the flow rate and pressure across the senso...

Страница 5: ...ation For connection of the sensor to the electronics a 25 cable is supplied An additional 100 feet of interconnect cable may be added using 07 0100 junction box All required spare parts are also prov...

Страница 6: ...Selectable PID controller on main analog output PID controller can operate with instrument configured as loop power transmitter or as one of the two outputs on the AC powered instrument PID includes...

Страница 7: ...with sensor flowcell Ambient Temperature Analyzer Service 20 to 60 C 4 to 140 F Sensor Service 5 to 55 C 23 to 131 F Storage 5 to 70 C 22 to 158 F Ambient Humidity 0 to 95 non condensing EMI RFI Infl...

Страница 8: ...ptional Relays Three additional non isolated low power relays available SPST 1 Amp 30 VDC Analog Outputs Two 4 20 mA outputs Output one programmable for PPM peroxide or PID Output 2 programmable for P...

Страница 9: ...instrument in close proximity to the point of sensor installation this will allow easy access during calibration The sensor to instrument distance should not exceed 100 feet To maximize signal to nois...

Страница 10: ...instrument is then attached to the wall using the four outer mounting holes in the bracket These holes are slotted to accommodate two sizes of u bolt that may be used to pipe mount the unit Slots wil...

Страница 11: ...ATI Q46 84 Hydrogen Peroxide Monitor Part 2 Analyzer Mounting 11 O M Manual Rev H 6 17 Figure 4 Wall Mounting Diagram Figure 5 Pipe Mounting Diagram...

Страница 12: ...tems The panel mounting bracket kit must be ordered separately part 05 0068 This kit contains a metal retainer bracket that attaches to the rear of the enclosure 4 screws for attachment of this bracke...

Страница 13: ...w concentrations below 0 5 PPM this method should always be used Mechanical installation of the flow cell requires that it be mounted to a wall or other convenient flat surface Alternatively the mount...

Страница 14: ...let flow is well controlled can use a simpler sealed flowcell Using this flowcell requires that flow be controlled externally to about 400 cc min Variable flow rate or variable pressure will cause uns...

Страница 15: ...relevant local codes and instructions contained in this operating manual Observe the analyzer s technical specifications and input ratings Proper electrical disconnection means must be provided prior...

Страница 16: ...the microprocessor assembly located in the front section of the enclosure This cable may be unplugged from the front section of the monitor if service is needed but should normally be left in place d...

Страница 17: ...e proper power to the contacts For applications that require the same switched operating voltage as the Q46 115 or 230 V power may be jumpered from the power input terminals at TB7 Relay wiring is con...

Страница 18: ...o connect to the optional 3 relay card Figures 11 OR the optional third analog output Out 3 Figure 12 The Q46 can be configured for only one of these optional features and the hardware for either opti...

Страница 19: ...removed easily from the flowcell Cord grips used for sealing the cable should be snugly tightened after electrical connections have been made to prevent moisture incursion When stripping cables leave...

Страница 20: ...o available to provide a break point for long sensor cable runs Route signal cable away from AC power lines adjustable frequency drives motors or other noisy electrical signal lines Do not run sensor...

Страница 21: ...ax 100 feet a junction box must be used The junction box is shown in Figure 15 and is supplied with a conduit hub on one end and a sensor cable gland on the other end NOTE If sensor is experiencing Lo...

Страница 22: ...2O2 Sensor Assembly Follow the procedure below to prepare the sensor for operation CAUTION Use extreme care when filling the H2O2 sensor with electrolyte The electrolyte is very acidic and can cause e...

Страница 23: ...lowly screw the chamber onto the sensor body A small amount of electrolyte will run out of the hole from which the vent screw was removed Place a paper towel around the sensor to absorb the electrolyt...

Страница 24: ...from this user interface no internal jumpers pots etc Figure 17 User Interface MENU ICONS UNITS 12 CHARACTER SECONDARY DISPLAY MEMBRANE KEYPAD MENU ESC ENTER A B DIAG FAIL HOLD CAL CONF MENU ICONS UN...

Страница 25: ...tem for change and to store any change 6 12 Display The large custom display provides clear information for general measurement use and user configuration There are three main areas of the display the...

Страница 26: ...the output is locked to the last input value measured when the HOLD function was entered HOLD values are retained even if the unit power is cycled FAIL The FAIL icon indicates that the system diagnost...

Страница 27: ...ompleted pressing ENTER again stores the variable and stops the flashing Pressing ESC aborts the change and also exits user entry mode The starting default screen is always the Measure Menu The UP arr...

Страница 28: ...AIL Relay B CON Set Default Setpnt A Delay A Phase A Hyst A Relay A AL Relay A FAIL Relay A CON Setpnt A HI Delay A HI Setpnt A LO Hyst A HI Delay A LO Phase A Hyst A LO Phase B Slope CONTROL Fail Val...

Страница 29: ...er can be placed in manual while viewing this screen by pressing and holding the ENTER key for 5 seconds until a small flashing m appears on the screen At that point the controller output can be adjus...

Страница 30: ...ensitivity is to specification in any user selected range 6 24 Configuration Menu CONFIG The Configuration Menu contains all of the general user settings Entry Lock This function allows the user to lo...

Страница 31: ...alue will flash Use the UP arrow key to modify the desired display value Press ENTER to store the new value Com Mode Sets digital communication mode of analyzer Optional digital communication card mus...

Страница 32: ...B and C can be used in two ways as a setpoint control or as an Relay C Mode alarm The two settings for Relay B Mode are CON and FAIL The CON setting enables normal setpoint operation for Relay B C Re...

Страница 33: ...pting to maintain by Iout1 PID adjusting output value It is the nature of the PID controller that it never actually gets to the exact value and stops The controller is continually making smaller and s...

Страница 34: ...tions set the optional third 4 mA and 20 mA Set 20 mA 3 current loop output points for the analyzer The output may be set to temp H2O2 track temperature default or peroxide The values stored for the 4...

Страница 35: ...LO the relay energizes and the LCD indicator illuminates when the H2O2 level drops below the setpoint The failsafe setting does have an impact on this logic The description here assumes the failsafe s...

Страница 36: ...st automatically If Relay C Mode is set to CON see Relay C Mode then Relay C will function identically to Relay A Relay C settings appear in the CONFIG menu list automatically If Relay D E or F Mode i...

Страница 37: ...tion on Press the ESC key to go to the DIAG Menu and scroll to Sim Output using the UP arrow key Press ENTER Follow the instructions under Sim Output see following page CAUTION There is no time out on...

Страница 38: ...ontrol of the SIM screen until the ESC key is pressed Note If the HOLD function is engaged before the Sim Output function is engaged the simulated output will remain the same even when the ESC key is...

Страница 39: ...F under this normal condition Therefore when the trip condition occurs the closed N O contacts will be opened relay de energized and the LCD icon will illuminate In addition a power fail would also ca...

Страница 40: ...d need not be distilled but it must not contain hydrogen peroxide If the sensor is being used for the first time allow the sensor to sit undisturbed for at least 4 hours If the sensor has been in use...

Страница 41: ...the sensor into the flow cell until the membrane cap bottoms out on the acrylic flow cell Do not over tighten 2 Turn on the inlet water flow to the flow cell and adjust the inlet flow rate so that wa...

Страница 42: ...ibrium with the solution in order to provide the highest accuracy When moving the sensor between widely different temperature conditions it may be necessary to allow the sensor to stabilize as much as...

Страница 43: ...chemicals etc that enable the controller to add or subtract chemical rapidly enough This is not only specific to pumps and valves but also to line sizes delays in the system etc This section is includ...

Страница 44: ...ber between 3 5 to start 1 2 more than P Like proportional gain increasing the integral term will cause the controller action to be quicker D Derivative gain The addition of derivative control can be...

Страница 45: ...and hold the ENTER key for about 5 seconds You will see a small m show up between the value and the mA value This indicates you are now in manual mode Once in manual you may increase the PID output by...

Страница 46: ...al at the extreme ranges of the measuring scale In addition if a chemical process is not symmetrical that means it responds differentially to the addition and subtraction of chemical It is important i...

Страница 47: ...ifficult to clean effectively Immersing the tip of the sensor in 1N nitric acid solution will sometimes remove deposits that cause low sensitivity but this is not always the case The recommended pract...

Страница 48: ...f the sensor to be submerged Allow the electrodes to soak in distilled water for 30 minutes 4 Put a new membrane and fresh electrolyte in the electrolyte chamber and reassemble the sensor Connect to t...

Страница 49: ...4 20 mA outputs do not exceed the limits in the Instrument Specifications During troubleshooting it is many times helpful to disconnect all these outputs and place wire shorts across the terminals in...

Страница 50: ...monitor to verify that the millivolt value is actually 400 mV 4 Check TC drive circuit Place a wire short between the GREEN and BLACK terminals With a DVM measure the voltage between terminals BLACK a...

Страница 51: ...above TC Error TC may be open or shorted Check sensor wiring and perform RTD test as described in sensor manual Check j box connections Figure 22 Display Messages MESSAGE DESCRIPTION POSSIBLE CORRECTI...

Страница 52: ...e red and white wires The red wire is part of the RTD circuit and the white wire is part of the measuring cell There should be no connection Reading resistance you should find an open circuit value of...

Страница 53: ...C with 3 rd 4 20mA output 03 0409 Q46 P S Assy 100 240 VAC with 3 relay exp Board 03 0410 Q46 P S Assy 12 24 VDC 03 0411 Q46 P S Assy 12 24 VDC with 3 rd 4 20mA output 03 0412 Q46 P S Assy 12 24 VDC w...

Страница 54: ...of extended exposure to excessive gas concentrations Should inspection indicate that sensors have been expended rather than failed prematurely the warranty shall not apply The Manufacturer assumes no...

Страница 55: ...DUCTS NH3 Ammonia CO Carbon Monoxide H2 Hydrogen NO Nitric Oxide O2 Oxygen CO Cl2 Phosgene Br2 Bromine Cl2 Chlorine ClO2 Chlorine Dioxide F2 Fluorine I2 Iodine HX Acid Gases C2H4O Ethylene Oxide C2H6O...

Отзывы: