Q45H Training Handout 4.05 

Page 3 of 11

 

Chlorine Calibration 

 

Once  power  is  applied  and  sample  water  is  running  past  the  sensor  (via  the  flow  cell)  a  good  6-8  hours  should  be 
allowed for complete polarization ‘of a brand new sensor’ (unless using the polarizer as above) re-builds after this time 
should only take about 15 – 20 minutes.  It is generally unnecessary to set the zero at start up and/or every calibration; 
however,  it  should  be  checked  in  the  MEASURE  display,  go  UP  to  see  zero  =  0.00  an  (if  greater  than  +/-25nA,  an 
electronic zero 

cannot 

be carried out) see 

chlorine zero cal

 below.  

 

There are two options in the chlorine calibration menu, “1 Ref” and “2 Zer”  

 

Ref 

Used to set the monitor to read the same as manual sample (e.g. Palin or Hach test with DPD) during 
calibration 
 

Zer 

Used to check the zero offset of the sensor. Zero offsets on ATi sensor are very stable and it is not 
necessary to re zero the sensors before every calibration 
 

 

 
 

Chlorine Zero Cal 

“2 Zer” 

Chlorine sensors have extremely low offset currents at zero.  For this reason, it is normally sufficient to simply leave the 
zero at the factory default of 0.0 nA.  As an alternative, an electronic zero can be set by disconnecting the sensor from 
the cable.  
 
The steps below assume that the sensor has been prepared in accordance with page 2, Chlorine Sensor Assembly. Note 
that the 8 hour waiting time in step 2 below is not required if the monitor has been running for 8 hours prior to power 
and water sample flowing.  If the unit has been running with the sensor connected, the sensor will normally return to a 
stable zero within 15 minutes. 

 

1.

 

If the sensor requires zeroing simply go to the MEASURE mode, press the UP arrow until it reads ‘zero = ? n/A  

 
2.

 

Now disconnect the chlorine sensor and watch the readings drop to zero (leave for a couple of minutes to allow a 

stable zero. 

 
3.

 

Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.  

Cal Chlor

 will then be 

displayed. 

 
4.

 

Press the ENTER key.  The screen will display a flashing 

1-Ref

 for span calibration or a 

2-Zer

 for zero calibration.  

Using the UP arrow key, set for a 

2-Zer

 zero calibration and press ENTER. 

 

5.

 

The system now begins acquiring data for the sensor zero calibration value.  As data is gathered, the units for sensor 
current in nanoamps (nA) and temperature may flash.  Flashing units indicate that this parameter is unstable.  The 
calibration data point acquisition will stop only when the data remains stable for a pre-determined amount of time.  
This can be overridden by pressing ENTER.  If the data remains unstable for 10 minutes, the calibration will fail and 
the message 

Cal

 

Unstable

 will be displayed. 

 
If accepted, the screen will display the message 

PASS

 with the new sensor zero reading (offset), then it will return 

to  the  main  measurement  display.    If  the  calibration  fails,  a  message  indicating  the  cause  of  the  failure  will  be 
displayed and the FAIL icon will be turned on.  The range of acceptable value for sensor offset is -25 nA to +25 nA.  
Should a FAIL occur, carefully inspect the sensor for a tear in the membrane.  It will probably be necessary to rebuild 
the  sensor  as  described  on  page  2,  Chlorine  Sensor  Assembly.    Should  the  offset  value  remain  high  and  result  in 
calibration  failures, review  the Service section  of this  manual, and then  contact  the service dept. at ATI for further 
assistance. 

 

The sensor zero offset value in nA from the last zero calibration is displayed on the lower line of the Default Menus 
for information purposes.   

 

Содержание Q45H/62

Страница 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Страница 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Страница 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Страница 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Страница 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Страница 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Страница 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Страница 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Страница 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Страница 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Страница 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...

Отзывы: