ATI Technologies Q45H/62 Скачать руководство пользователя страница 3

 

Q45H Training Handout 

Page 2 of 11

 

Chlorine Sensor Preparation 

 

The chlorine sensor supplied with the Q45H is shipped dry.  It will not operate until it is prepared by adding 
electrolyte and a membrane.  Preparation of the sensor for operation must be done carefully.  A qualified technician 
should do the procedure and it should only be done when the system is ready for operation. Until then, it is best to 
leave the sensor in the condition in which it is received. 

 
 

 

 

Follow the procedure below to prepare the chlorine sensor for operation: 

 

1.

 

Unscrew the electrolyte chamber from the assembled sensor and also remove the fill screw from the side of the 
sensor body. 

 
2.

 

Remove the silver membrane cap from the bottom of the chamber and discard the protective membrane.  O-rings are 
contained in grooves on both the bottom and top of the chamber.  Be sure that these o-rings remain in place. 

 
3.

 

From the package of membranes supplied with the sensor, place a new membrane into the membrane holder. The 
membrane can go in either side. Please try not to touch the centre of the membrane as greasy fingers can block the 
micro-porous holes.  

The membrane is white in colour and is separated from other membranes by a light 

blue paper spacer.

 

 
4.

 

Screw the silver membrane cap on to the chamber until you feel the o-ring compress.  Hand tight compression is all 
that is needed.  Do not use tools to tighten.  The membrane should be flat across the bottom of the chamber without 
wrinkles. 

 
5.

 

Fill the chamber with electrolyte until the level reaches the bottom of the internal threads. Add a drop more to be 
sure that you have filled it fully. 

 
6.

 

Unscrew the small screw near the base of the sensor. Be careful not to lose the O ring attached. Slowly screw the 
chamber onto the sensor body.  A small amount of electrolyte will run out of the hole from which the fill screw was 
removed.  The electrolyte is harmless and will not irritate skin. IF NO ELECTROLYTE LEAKS, YOU MAY NOT HAVE PUT 
SUFFICIENT IN THE ELECTROLYTE CHAMBER AND YOU MUST DO THIS AGAIN. Tighten the chamber until the o-ring 
at the top of the chamber is compressed.  Once again, do not use tools to tighten. 

 
7.

 

Shake excess electrolyte from the fill hole on the side of the sensor and replace the fill screw. 

 

The membrane should be stretched tightly across the tip of the sensor.  The sensor is now ready for connection to the 
electronics. It should be left connected and either in the flow cell with chlorinated water or in beaker of water for at least 
8 hours prior to operation.  

A polarizer can be used to dramatically reduce the length of time required on start up. Build 

the sensor as above, place some wet tissue into the  red cap, push onto sensor  end, push onto the polarizer and store 
(remember to replace the membrane when using > 2 months after due to particulate forming on the membrane from the 
wet tissue/material used. 

 

Содержание Q45H/62

Страница 1: ...ee Chlorine Measurement System Training Handout European Office Analytical Technology Unit 1 2 Gatehead Business Park Delph New Road Delph Saddleworth OL3 5DE Tel 01457 873 318 Fax 01457 874 468 Email...

Страница 2: ...the escape key aborts all changes to the current screen and backs the user out one level in the software tree The manual will refer to this key as either MENU or ESC depending upon its particular fun...

Страница 3: ...ssion is all that is needed Do not use tools to tighten The membrane should be flat across the bottom of the chamber without wrinkles 5 Fill the chamber with electrolyte until the level reaches the bo...

Страница 4: ...in 15 minutes 1 If the sensor requires zeroing simply go to the MEASURE mode press the UP arrow until it reads zero n A 2 Now disconnect the chlorine sensor and watch the readings drop to zero leave f...

Страница 5: ...nd press ENTER AT THIS POINT CARRY OUT THE DPD TEST AND WAIT FOR AT LEAST 20 SECONDS PER DPD TABLET AND CARRY OUT 2 TO 3 TO GET THE CORRECT AVERAGE 4 The system now begins acquiring data for the calib...

Страница 6: ...r calibration this minimum is required due to potential error of dpd test pH Calibration Must do first Keys to Press Screen Display User Action Required 1 MENU ESC until CAL MENU 2 ENTER Cal Cl2 3 Cal...

Страница 7: ...CONFIGURATION and program in CONTROL menu There are 2 x 4 20mA outputs and 2 x relays can be configured for alarm control or fail Initial set up in CONFIGURATION and then program in CONTROL menu The...

Страница 8: ...and check that the o rings are in place and are not broken If after time this happens again it may be due to micro cavitation created by the pump you are using drawing the electrolyte out of the memb...

Страница 9: ...d The difference between the very experienced and less experienced user is that the experienced user would calibrate the colorimeter prior to calibration take 3 tests and average them out The less exp...

Страница 10: ...the fill screw from the side of the sensor body 2 Remove the front nut from the bottom of the chamber and discard the protective membrane O rings are contained in grooves on both the bottom and top o...

Страница 11: ...Q45H Training Handout Page 10 of 11 Q45H 62 OPERATIONAL LOG PRINT OFF AND USE Date Q45H pH Value Portable pH Value Q45H CL2 Value DPD CL2 Value Pre Cal Slope Post Cal Slope Maintenance...

Отзывы: