The transmitter records gas concentrations in one of 12 intervals ranging from 1 to 60 minutes, providing
data from 11 to 474 days. Table 17 details the sampling intervals, and the samples/day and totals days for
each interval.
Table 17. Data-log sampling metrics
Sampling
(Minutes)
Samples/Day
Total
Days
1
1440
11
2
720
22
3
480
32
4
360
43
5
288
54
6
240
64
10
144
104
12
120
124
15
96
152
20
72
196
30
48
278
60
24
474
The gas concentration reading is recorded as an instantaneous value and is not averaged or filtered in any
way. When the data log memory is filled, new records overwrite older ones.
Data Log Menu
The Data Log Menu permits access to configuration, review, and print menus.
Figure 59. Data Log Menu
Item
Select to …
Setup
Configure the data log settings (see Data Log Setup Menu below).
View
View the logged data as a graph or single text records.
Send a tabular ASCII report to the device connected to the COM port
(see
on the next page). Note: only available on
transmitters with an RS232 or RS485 interface. The data log must not
be empty, and the COM protocol must be set to ASCII. Otherwise,
the transmitter will display an exception message.