ATI Technologies A15/82 Скачать руководство пользователя страница 24

ATI 

Model A15/82 Fluoride Monitor with Auto-Calibration

 

 

 

23 

O&M Manual 
Rev-G (6/10) 

OPERATION 

 
 

With the chemistry module in operation and power turned on to the monitor, the system should now be 

displaying a fluoride concentration from the sample.  The value measured initially will probably be fairly close to 
the actual value, but the monitor is not factory calibrated.  While the monitor contains an automatic calibration 
feature, a manual 2-point calibration must be done at startup.   
 

The calibration procedure will require preparation of two fluoride standards.  A 500 cc bottle of 100 PPM 

fluoride standard is supplied with the monitor system.  Standards of 1 PPM and 10 PPM should be prepared for 
use in calibration.  Two bottles are supplied for holding the mixed standards.  You will also need a pipette and a 
volumetric cylinder for mixing your standards.   
 

For mixing the 1 PPM standard, pipette 1 cc. of the 100 PPM standard into a 100 cc volumetric flask.  

Dilute to 100 cc using either unfluoridated drinking water or distilled water.  Using unfluoridated water from the 
system has the advantage of producing a standard with the same background composition of the water you will 
be monitoring.  If this is not feasible, distilled water can be used instead as there will be little real difference in the 
end  result.    Pour  the  mixed  standard  into  the  1  PPM  standards  bottle  and  then rinse the volumetric flask with 
distilled water. 
 

For the 10 PPM standard, pipette 10 cc. of the 100 PPM standard into the volumetric flask and dilute to 

100 cc. as before.  Pour this standard into the bottle marked 10 PPM.  Keep these two standards bottles capped 
for use in calibration checks as needed. 
 

CALIBRATION  

 
 

The fluoride monitor electronics is designed to allow up to a 5 point calibration for ion selective sensors.  

For  fluoride  measurement,  only a two point calibration is necessary.  The standards are fed into the chemistry 
module  using  an  extension  tube  supplied  with  the  unit.    Remove  the  sample  pickup  tube  from  the  overflow 
chamber  and  connect  the  extension  tube.    A  barb  fitting  is  supplied  on  the  extension  tube  for  this  purpose.  
Follow the procedure below. 

 
 

1.  Place  the  extended  pickup  tube  into  the  1  PPM  standard  mixed  previously  and  allow  the  standard  to  be 

pumped into the chemistry module.     

 
2.  Press the MODE key on the monitor once and the display will indicate “ppm  X point cal”.  Press the CAL key 

once and the display will read “CAL POINT Nr. 1”.  Press CAL again and the display will read “CAL P1:  X.X 
mV”.  This display shows the current mv. signal from the fluoride sensor (this value will normally be around 
+40 mV for the glass electrode and -350 mV for the polymer sensor.)   

 
3.  After  the  1  PPM  standard  has  been  flowing  for  5  minutes,  the  mv.  signal  should be stable to within a few 

tenths  of  a  millivolt.    Press  ENTER  and  the  display  will  indicate  “P1  DECADE:  100.0”.    Press  the  DOWN 
arrow and the display will read “P1 VALUE:  10.00”.  Press ENTER and then use the DOWN arrow to adjust 
the value to 1.00 (hold button down and display will scroll down).  Then press ENTER. 

 
4.  The  display  will  now  indicate  “CAL  POINT  Nr.  2”.    Remove  the  standards  pickup  tube  from  the  1PPM 

standards bottle and place it into the 10 PPM standards bottle.  Allow the system to run for about 5 minutes 
on the 10 PPM standard.   

 
5.  Press  CAL  and  the  display  will  read  “CAL  P2:    X.X  PPM”.    As  before,  the  millivolt  value  should  be  fairly 

stable after the 10 PPM standard has been flowing for 5 minutes (this value will normally be around -15 mV 

Содержание A15/82

Страница 1: ...UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Saddleworth OL3 5DE Ph 800 959 0299 Ph 0800 018 4020 610 917 0991 44 0 1457 873 318 Fax 610 91...

Страница 2: ...exposure to excessive gas concentrations Should inspection indicate that sensors have been expended rather than failed prematurely the warranty shall not apply The Manufacturer assumes no liability f...

Страница 3: ...CONNECTION 13 ELECTRICAL CONNECTIONS 14 START UP 16 FRONT PANEL LOCK 17 CONFIGURATION 18 CHEMISTRY MODULE STARTUP 21 SENSOR INSTALLATION 21 FLOW STARTUP 21 AUTOMATIC CALIBRATION VALVE 22 OPERATION 23...

Страница 4: ...8 Figure 3 NEMA 4X Wall Mount Dim 9 Figure 4 Fluroide Montior Chem Module 10 Figure 5 Sample Inlet and Drain Chamber 11 Figure 6 Buffer Bottle Holder Dim 12 Figure 7 Reagent Pickup Tubing Assy 13 Figu...

Страница 5: ...pare Parts Kit Quantity 1 Fluoride Buffer Powder in gallon container Quantity 2 Buffer Calibration Standard bottle holder not supplied if stainless mounting plate is ordered Quantity 1 Operation Maint...

Страница 6: ...entration a small amount of buffer is metered into the sample to provide a constant background condition With other parameters held constant fluoride activity and fluoride concentration are closely re...

Страница 7: ...ATI Model A15 82 Fluoride Monitor with Auto Calibration 6 O M Manual Rev G 6 10 Figure 1 Fluoride Monitoring System...

Страница 8: ...600 ohm maximum load Programmable output span Output may be inverted Operating Conditions 0 50 C 0 95 R H non condensing Power 110 220 VAC 10 50 60 Hz Enclosure Panel mount standard NEMA 4X PPO wall m...

Страница 9: ...feet of the chemistry module due to the limited sensor cable length If the optional stainless steel mounting plate was purchased all components mount to this plate with the plate mounted to a wall If...

Страница 10: ...t that bolts directly to a wall or other flat surface The enclosure is supplied with special mounting bracket for convenience in mounting the enclosure A hinged door on the front of the enclosure prov...

Страница 11: ...components required to condition the sample and measure the fluoride concentration under stable conditions This package includes sample and buffer pumps the fluoride ISE and a sensor flow chamber Fig...

Страница 12: ...ow assembly on its aluminum mounting bracket The bracket is supplied to hold the inlet assembly under the sample pump and to allow waste from the chemistry module to flow back into the drain chamber T...

Страница 13: ...chemistry module The brackets are designed for wall mounting and will hold one gallon plastic containers When mounting these brackets be sure to leave enough room between the bottom of the chemistry m...

Страница 14: ...t After the buffer solution is ready feed the end of the tubing opposite the filter up through the ceramic tube Then insert the tube assembly into the solution and press the rubber stopper in place Wh...

Страница 15: ...a center conductor and a shield Refer to Figure 8 for a detail of the terminal connections on the rear of the fluoride monitor 1 AC POWER NEUTRAL 14 4 20 mA OUTPUT CHANNEL 1 2 AC POWER 110V HOT 15 NO...

Страница 16: ...cord is supplied Permanent wiring should be made to the terminals shown in Figure 9 below The sensor wire is an unbroken coaxial wire running directly into the monitor In addition a separate 5 conduc...

Страница 17: ...ay A as marked on the back of the monitor An open square o indicates a deactivated relay while a solid square n indicates an activated relay More information on the control relays is provided in a lat...

Страница 18: ...ed in order to prevent unauthorized adjustments to the instrument When the front panel is locked the MODE key will still allow the display to be stepped through the alternate displays but will not all...

Страница 19: ...the overall display range 10 00 is the default but ranges of 100 0 and 1000 can be selected if desired Select using the UP or DOWN keys and then press ENTER 4 LARGE S RT 2 0s This routine allows for...

Страница 20: ...where the A relay is being used for control since it provides an alarm if the control function does not maintain the desired value for some period of time 12 AL SET B OFF This routine is similar to 8...

Страница 21: ...stems However the value can be adjusted if a different standard is used 23 ZERO MAX 10 mV When the automatic calibration is complete the monitor analyzes the zero offset of the sensor compared with a...

Страница 22: ...t sample flow using the flow control valve installed on the sample line Adjust the flow so that water is overflowing from the inlet chamber into the drain chamber The rate at which water overflows thi...

Страница 23: ...switch calibration standard into the measuring cell when called for by the monitor Tubing from the calibration standard is routed through the bottom section of the pinch valve normally closed and the...

Страница 24: ...for use in calibration checks as needed CALIBRATION The fluoride monitor electronics is designed to allow up to a 5 point calibration for ion selective sensors For fluoride measurement only a two poin...

Страница 25: ...ly Press ENTER 6 The display will change to CAL POINT Nr 3 Press ENTER and the display will indicate END POINT CAL Press ENTER and the calibration will be updated After completion of the calibration d...

Страница 26: ...ng no change in the default timing for calibration and then will switch to If the calibration should fail due to excessive drift of the sensor which could be caused by buffer feed failure calibration...

Страница 27: ...and pour this into the gallon bottle Fill the gallon container to about the bottom of the neck with the water poured into the separate container This solution will be 1 00 to 1 05 PPM fluoride and can...

Страница 28: ...e MODE key until SA appears on the left side of the display The full display will indicate SA X XXppm oLO This refers to setpoint A and indicates the current setting for control relay A X XX is the cu...

Страница 29: ...m To program the Alarm Relay for different low and high setpoints proceed as follows 1 Press the MODE key until the display show AL at the left side The numbers shown to the right of AL are the curren...

Страница 30: ...standard for about 5 minutes Should the fluoride system be shut down for a length of time it will be nessary to remove the sensor rince the sensor with de ionized water and replace the protective cap...

Страница 31: ...the inlet fitting and the outlet tubing from the outlet side of the pump Remove the tubing clamp plate and push the tube holders out of their slots Push on the tubing from inside the retaining block...

Страница 32: ...ube Holder 1 channel Sample Pump 54 0008 Tube Holder 2 channel Reagent Pump 54 0009 Pressure plate 54 0010 Pump motor for sample pump 120 VAC 54 0011 Pump motor for buffer pump 120 VAC 54 0012 Pump mo...

Отзывы: