Model A15/82 Fluoride Monitor with Auto-Calibration

 

 

 

16 

O&M Manual 
Rev-G (6/10) 

    START-UP 

 
 

Prior  to  operation,  recheck  electrical  connections  to  be  sure  everything  is  in  accordance  with  Figure  8 

and Figure 9.  If everything is in order, power may be applied.  When power is turned on, the display will come 
up with a program and version number, which will remain on the display for a few seconds.  The display will then 
change  to  a  display  of  the  fluoride  value  and  an  indication  of  the  status  of  control  relays.    This  is  the  main 
operating display, and should look like the display below: 
 

D1

 

XX.XX ppm   

o

A  

 

MAIN DISPLAY 

 

The XX.XX portion will be numbers.  The initial values could be almost anything, depending on whether 

the  sensor  has  been  connected  and  what  the  condition  of  the  sensor  is.    The  values  are  not  important  at  this 
point.    The  display  may  indicate  arrows  to  the  left  or  right,  but  this  should  not  be  of  concern  until  sample  is 
flowing.  The next portion of the display indicates control relay status.  The letter A refers to control relay A as 
marked on the back of the monitor.  An open square (

o

) indicates a deactivated relay while a solid square (

n

indicates an activated relay.  More information on the control relays is provided in a later section of this manual.  
Note  that  if  the  automatic  calibration  feature  of  this  system  is  disabled,  relay  B  status  will  also  be  displayed.  
Beyond  the  main  display,  a  series  of  alternate  displays  are  available  to  the  operator.    These  displays  may  be 
selected by pressing the MODE key on the front panel.  It is a good idea to step through the displays to become 
familiar  with  each  one.    Many  of  these  alternate  displays  serve  as  a  starting  point  for  programming  functions 
described later in this manual.  The alternate display sequence is as follows: 
 
 

D2

 

ppm     2 point cal 

 NUMBER OF CALIBRATION POINTS  

 

D3

 

   XX.X mV        X

-

 

FLUORIDE ISE VOLTAGE OUTPUT 

 

D4

 

TEMP:     20.0

°

 C

 

SAMPLE TEMPERATURE DISPLAY 

 

D5

 

THERMOC.    ON 

TEMPERATURE COMPENSATION OFF 

 

D6

 

SA   X.XXPPM   

o

L

 

RELAY A SETPOINT AND FUNCTION 

 

D7

 

AL X.XX/X.XX    F

-

 

ALARM RELAY SETPOINTS 

 

D8

 

CLEANING:  OFF 

CLEANING FUNCTION TURNED OFF 

 

D9

 

AUTO  CAL 

AUTOMATIC CALIBRATION ENABLED 

 

D10

 

O1   4.0mA/ 0.00 F

-

 

ANALOG #1 OUTPUT INDICATION 

 

D11

 

CONFIGURATION 

ACCESS TO SETUP ROUTINES 

 

D12

 

IC7685.010     R2.20 

PROGRAM AND RELEASE NUMBER 

Содержание A15/82

Страница 1: ...UK Limited 6 Iron Bridge Drive Unit 1 2 Gatehead Business Park Collegeville PA 19426 Delph New Road Delph Saddleworth OL3 5DE Ph 800 959 0299 Ph 0800 018 4020 610 917 0991 44 0 1457 873 318 Fax 610 91...

Страница 2: ...exposure to excessive gas concentrations Should inspection indicate that sensors have been expended rather than failed prematurely the warranty shall not apply The Manufacturer assumes no liability f...

Страница 3: ...CONNECTION 13 ELECTRICAL CONNECTIONS 14 START UP 16 FRONT PANEL LOCK 17 CONFIGURATION 18 CHEMISTRY MODULE STARTUP 21 SENSOR INSTALLATION 21 FLOW STARTUP 21 AUTOMATIC CALIBRATION VALVE 22 OPERATION 23...

Страница 4: ...8 Figure 3 NEMA 4X Wall Mount Dim 9 Figure 4 Fluroide Montior Chem Module 10 Figure 5 Sample Inlet and Drain Chamber 11 Figure 6 Buffer Bottle Holder Dim 12 Figure 7 Reagent Pickup Tubing Assy 13 Figu...

Страница 5: ...pare Parts Kit Quantity 1 Fluoride Buffer Powder in gallon container Quantity 2 Buffer Calibration Standard bottle holder not supplied if stainless mounting plate is ordered Quantity 1 Operation Maint...

Страница 6: ...entration a small amount of buffer is metered into the sample to provide a constant background condition With other parameters held constant fluoride activity and fluoride concentration are closely re...

Страница 7: ...ATI Model A15 82 Fluoride Monitor with Auto Calibration 6 O M Manual Rev G 6 10 Figure 1 Fluoride Monitoring System...

Страница 8: ...600 ohm maximum load Programmable output span Output may be inverted Operating Conditions 0 50 C 0 95 R H non condensing Power 110 220 VAC 10 50 60 Hz Enclosure Panel mount standard NEMA 4X PPO wall m...

Страница 9: ...feet of the chemistry module due to the limited sensor cable length If the optional stainless steel mounting plate was purchased all components mount to this plate with the plate mounted to a wall If...

Страница 10: ...t that bolts directly to a wall or other flat surface The enclosure is supplied with special mounting bracket for convenience in mounting the enclosure A hinged door on the front of the enclosure prov...

Страница 11: ...components required to condition the sample and measure the fluoride concentration under stable conditions This package includes sample and buffer pumps the fluoride ISE and a sensor flow chamber Fig...

Страница 12: ...ow assembly on its aluminum mounting bracket The bracket is supplied to hold the inlet assembly under the sample pump and to allow waste from the chemistry module to flow back into the drain chamber T...

Страница 13: ...chemistry module The brackets are designed for wall mounting and will hold one gallon plastic containers When mounting these brackets be sure to leave enough room between the bottom of the chemistry m...

Страница 14: ...t After the buffer solution is ready feed the end of the tubing opposite the filter up through the ceramic tube Then insert the tube assembly into the solution and press the rubber stopper in place Wh...

Страница 15: ...a center conductor and a shield Refer to Figure 8 for a detail of the terminal connections on the rear of the fluoride monitor 1 AC POWER NEUTRAL 14 4 20 mA OUTPUT CHANNEL 1 2 AC POWER 110V HOT 15 NO...

Страница 16: ...cord is supplied Permanent wiring should be made to the terminals shown in Figure 9 below The sensor wire is an unbroken coaxial wire running directly into the monitor In addition a separate 5 conduc...

Страница 17: ...ay A as marked on the back of the monitor An open square o indicates a deactivated relay while a solid square n indicates an activated relay More information on the control relays is provided in a lat...

Страница 18: ...ed in order to prevent unauthorized adjustments to the instrument When the front panel is locked the MODE key will still allow the display to be stepped through the alternate displays but will not all...

Страница 19: ...the overall display range 10 00 is the default but ranges of 100 0 and 1000 can be selected if desired Select using the UP or DOWN keys and then press ENTER 4 LARGE S RT 2 0s This routine allows for...

Страница 20: ...where the A relay is being used for control since it provides an alarm if the control function does not maintain the desired value for some period of time 12 AL SET B OFF This routine is similar to 8...

Страница 21: ...stems However the value can be adjusted if a different standard is used 23 ZERO MAX 10 mV When the automatic calibration is complete the monitor analyzes the zero offset of the sensor compared with a...

Страница 22: ...t sample flow using the flow control valve installed on the sample line Adjust the flow so that water is overflowing from the inlet chamber into the drain chamber The rate at which water overflows thi...

Страница 23: ...switch calibration standard into the measuring cell when called for by the monitor Tubing from the calibration standard is routed through the bottom section of the pinch valve normally closed and the...

Страница 24: ...for use in calibration checks as needed CALIBRATION The fluoride monitor electronics is designed to allow up to a 5 point calibration for ion selective sensors For fluoride measurement only a two poin...

Страница 25: ...ly Press ENTER 6 The display will change to CAL POINT Nr 3 Press ENTER and the display will indicate END POINT CAL Press ENTER and the calibration will be updated After completion of the calibration d...

Страница 26: ...ng no change in the default timing for calibration and then will switch to If the calibration should fail due to excessive drift of the sensor which could be caused by buffer feed failure calibration...

Страница 27: ...and pour this into the gallon bottle Fill the gallon container to about the bottom of the neck with the water poured into the separate container This solution will be 1 00 to 1 05 PPM fluoride and can...

Страница 28: ...e MODE key until SA appears on the left side of the display The full display will indicate SA X XXppm oLO This refers to setpoint A and indicates the current setting for control relay A X XX is the cu...

Страница 29: ...m To program the Alarm Relay for different low and high setpoints proceed as follows 1 Press the MODE key until the display show AL at the left side The numbers shown to the right of AL are the curren...

Страница 30: ...standard for about 5 minutes Should the fluoride system be shut down for a length of time it will be nessary to remove the sensor rince the sensor with de ionized water and replace the protective cap...

Страница 31: ...the inlet fitting and the outlet tubing from the outlet side of the pump Remove the tubing clamp plate and push the tube holders out of their slots Push on the tubing from inside the retaining block...

Страница 32: ...ube Holder 1 channel Sample Pump 54 0008 Tube Holder 2 channel Reagent Pump 54 0009 Pressure plate 54 0010 Pump motor for sample pump 120 VAC 54 0011 Pump motor for buffer pump 120 VAC 54 0012 Pump mo...

Отзывы: