Problem Possible 

Cause 

Correction 

Low Degree 

of Vacuum 

1.  Lack of oil 

2.  Oil is not clean 

3.  The oil inlet is blocked   

4.  The hoses or gas inlet are 

clogged 

5.  The pump is not suitable for 

your application 

1. Add oil to above the oil level line 

2. Change the oil 

3. Clean the oil inlet or clean the filter 

4. Check the connecting pipes 

5. Get suitable pump for your 

application 

Oil Leaks 

1.  The oil seal is damaged 

2.  The housing gasket is loose 

or worn out 

1. Change oil seal 

2. Change the housing gasket 

Oil Spray 

1.  Too much oil 

2.  The pressure at the gas inlet 

is too high or it has pumped 

too much 

1. Oil to the oil-level line 

2.

 

Change to a bigger pump 

Starting 

Difficulty 

1.  The oil temperature is too low 

2. Electrical 

malfunction 

3.  Foreign matter is in the pump 

1. Start the pump several times to try 

to heat the oil 

2. Check and have it fixed 

3. Check and remove it 

 

 

 

 

8. Maintenance 

·

Keep the pump clean and prevent foreign matter from entering. 

·

Keep the oil filled to the full level. 

Don’t let pump run without oil

·

Keep the oil clean. If the oil becomes dirty, muddy, or water or other volatile 

substance gets in, it will affect the performance of the pump and the oil should be 

replaced. Before replacing the oil, start the pump and have it run for about 30 minutes 

to make the oil thin. Stop the pump and drain the oil from the oil drain plug. At this 

time, open the gas inlet to release any pressure inside the pump. This will allow the 

oil to drain more freely. After making sure the pump is clean, put the drain plug back in 