Error 2
There is a reading „Err2“ on LCD display. Output value is about 3.8 mA. Measured value is below
lower limit of allowed full scale range. This state appears if measured temperature is lower than
approximately -210°C (i.e. low resistance of temperature sensor, probably short circuit).
Error 3
There is a reading „Err3“ on LCD display upper line.
Error of internal A/D converter appeared (converter does not respond, probably damage of A/D
converter). No measurements proceed. Output value is about 3.8 mA. It is a serious error, contact
distributor of the instrument.