English
21
- Remove the tube and place it on a non delicate and
easy to clean surface;
- Insert the tube in the relative tank and put it back in
the corresponding position:
• «Detergent»
• «Neutraliser»
• «Rinse aid».
- At the end, close the front door with the key.
- Start the program «Fill DOS1» for the detergent or
«Fill DOS2» or «DOS3» for the neutraliser or rinse aid.
Be careful not to reverse the position of the liquids.
Failure to follow these recommendations may damage
the thermodisinfector.
- With regard to the maximum dosage for each
program, follow the instructions of the chemical
manufacturer.
- To ensure efficiency of the dosing system, it is
necessary to regularly carry out the maintenance
operations set out in the «Maintenance» chapter.
- The indications concerning the storage and disposal
of chemical substances are provided by the respective
manufacturers and must be observed.
- Do not place the chemical containers on the
thermodisinfector.
Completely empty the thermodisinfector before
performing maintenance and before moving the
machine to avoid contact with the chemicals and to
protect the machine components.
5.6.2 Using and storing chemicals
Keep the containers tightly closed, stored in a dry place
and protected from the sun, out of reach of children
and outsiders. Possibly locked away. Optimal storage
temperature: from +0° to +25°C. The shelf life in the
original containers is indicated on the chemicals’ labels.
The manufacturer recommends a method for inventory
management (first in - first out).
The thermodisinfector can use up to three products
(two in the over-counter version) for dosing liquids.
The manufacturer recommends using cleaning agents
and chemical additives. The use of other products can
damage the machine.
The flow meters of chemicals are calibrated according
to the density of these tested products which ensure
correct operation.
The following combinations of process fluids have been
tested to verify the compatibility of the materials with
the components inside the device, for devices placed
on the market starting from 1 July 2019.
Liquids recommended:
Manufactu
-
rer
Detergent Neutraliser Rinse aid
Borer
Deconex
Prozyme
Alka-X
Deconex 64
Neutradry
Dr. Weigert
Neodisher
MediClean
Forte
Neodisher Z Neodisher
MediKlar
Dr. Weigert
Neodisher
MediClean
Dental
Neodisher Z
Dental
Neodisher
MediKlar
Dental
When chemical disinfectant levels are low, a warning
message appears on the display.
Chemical products dosing is set to an average value as
recommended by the manufacturer.
When the actual dosing of the chemical disinfectant
exceeds the tolerance of 5%, the system goes into
alarm.
Ö
If these recommended liquids are not used
in the machine, the flow meters must be
calibrated for the new liquids.
a
WARNING: Use only the chemical products
indicated in the table. The use of non-
certified products will void the warranty.
6 Using the machine
Before starting the machine, the operator in charge
must have read and understood this whole manual,
in particular the information given in section «3 Safety
and prevention».
Furthermore, before starting work, check that the
machine is in order and that all parts subject to wear
and deterioration are fully efficient.
6.1 Commissioning instructions
These control operations (reported below) are
performed to check if the machine works properly and
should be performed when the machine installation is
completed.
1) Open the tap that supplies water to the machine.
a
WARNING. The water should NOT flow
into the tank; otherwise the water loading
solenoid valves (in the lower compartment
of the machine) are dirty or blocked due to
long storage in the warehouse and therefore
they must be cleaned. Check that there are
no water leaks in the pipe fittings.