Programming Guide
©
Copyright Asyril S.A.
000.101.572
Version : A1
65/83
Address
(decimal)
Holding Register name
Read/
Write
Data Range
Comments
64+8
HR_MODBUS_SEQUENCE_STATUS
R
0: DONE
1: BUSY
2: ERROR
This Holding Register gives the status of the sequence execution trigged with
HR_SEQUENCE_EXECUTION_TRIG.
The transition DONE->BUSY occurs on the trigger rising edge if the data is
correct.
The transition DONE->ERROR occurs on the trigger rising edge if the data is
incorrect (e.g. index is out of range).
The transition BUSY->DONE occurs as soon as the sequence ends.
The transition BUSY->ERROR occurs if an Internal Modbus communication
error (See Section 6.4.2) before the sequence starts.
The transition ERROR->DONE can be controlled by setting the trigger to 0.
64+9
HR_SEQUENCE_REMAINING_TIME
R
0 to 32767
[ms]
This Holding Register gives the remaining time of the sequence execution.
During a sequence, the value decreases over time until it reaches 0.
Note: A value of 32767 means that the remaining time is greater than 32767ms.
64+10
HR_VIBRATION_SET_SELECTED
R
1 to 26
This Holding Register gives the current selected vibration set.
64+11
HR_SEQUENCE_SELECTED
R
1 to 26
This Holding Register gives the current selected sequence.
64+12
HR_BACKLIGHT_STATE_STATUS
R
0: OFF
1: ON
This Holding Register gives the current backlight state
64+13
HR_READ_WRITE_PARAMETER_READ_VALUE
R
0 to 65535
The parameter read trigged with HR_READ_WRITE_PARAMETER_TRIG
(Address 29) update this Holding Register value.
64+14
HR_FIRMWARE_VERSION_H
R
0 to 9
This Holding Register gives the first digit of the firmware version.
64+15
HR_FIRMWARE_VERSION_M
R
0 to 9
This Holding Register gives the second digit of the firmware version.
64+16
HR_FIRMWARE_VERSION_L
R
0 to 9
This Holding Register gives the third digit of the firmware version.
64+17
HR_USER_ACCESS_LEVEL
R
0: undefined
This Holding Register gives the current user access level. More information in