anufacturer’s declaration-electromagnetic immunity
The LU700 Series is intended for use in the electromagnetic environment (for professional healthcare) specified below. The customer or
the user of the LU700 Series
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment-
guidance (for
professional
healthcare
environment)
Electrostatic
discharge(ESD)
IEC
61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical
fast
transient/burst IEC 61000-
4-4
+ 2kV for power supply lines
+ 1kV for input/output lines
+ 2kV for power supply lines
Not applicable
Mains power quality should be that of a
typical
professional
healthcare
environment.
Surge IEC 61000-4-5
+ 0.5kV, +1kV line(s) to line(s)
+ 0.5kV, +1kV,+ 2kV line(s) to
earth
+ 0.5kV, +1kV line(s) to line(s)
Not applicable
Mains power quality should be that of a
typical
professional
healthcare
environment.
Voltage
Dips,
short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Voltage dips:
0
%
U
T
;
0,5
cycle
0 %
U
T
; 1 cycle
70 %
U
T
; 25/30 cycles
Voltage interruptions:
0 %
U
T
; 250/300 cycle
Voltage dips:
0
%
U
T
;
0,5
cycle
0 %
U
T
; 1 cycle
70 %
U
T
; 30 cycles
Voltage interruptions:
0 %
U
T
; 300 cycle
Mains power quality should be that of a
typical
professional
healthcare
environment. If the user of the
LU700C, LU700L
requires continued
operation
during
power
mains
interruptions, it is recommended that the
LU700C, LU700L be powered from an
uninterruptible power supply or a battery.
Power frequency(50, 60
Hz) magnetic field IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
60 Hz
The LU700C, LU700L power frequency
magnetic fields should be at levels
characteristic of a typical location in a
typical
professional
healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.