P67
Acoustic output reporting table
(EN IEC60601-2-37:2007+AMD1:2015, table 201.103)
Transducer Model: LU700L SN:LT702D-49-00007
Operating Model: B+CF Mode
Index Lebel
MI
TIS
TIB
TIC
At
surface
Below
surface
At
surface
Below
surface
Maximum Index Value
0.54
0.61
0.61
N/A
Index component value
B:0.56
CF:0.05
B:0.56
CF:0.05
N/A
B:0.56
CF:0.05
Associated
Parameter
P
r.a
at Z
MI
(MPa)
0.77
P
(mW)
B:22.23 CF:2.53
B:22.23 CF:2.53
N/A
P
1x1
(mW)
B:19.85 CF:2.26
B:19.85 CF:2.26
Z
S
(cm)
N/A
N/A
Z
b
(cm)
Z
MI
(cm)
0.50
Z
PII α
(cm)
0.50
f
awf
(MHz)
B:5.90
B:5.90 C:4.91
B:5.90 C:4.91
N/A
Other
Information
prr
(Hz)
8787.00
srr
(Hz)
7.21
n
pps
1
I
pa. α
at Z
PII α
(W/cm
2
)
64.83
-
I
spta. α
at Z
PII α
or Z
SII α
(mW/cm
2
)
21.3
I
spta.
at Z
PII
or Z
SII
(mW/cm
2
)
49.8
-
P
r.
at Z
PII
(MPa)
1.67
Operating
Control
Conditions
Display focus
(cm)
4.0
4.0
4.0
N/A
4.0
N/A
Display depth
(cm)
6.3
6.3
6.3
N/A
6.3
N/A
Working frequency
(MHz)
B:6.3
CF:5.0
B:6.3
CF:5.0
B:6.3
CF:5.0
N/A
B:6.3
CF:5.0
N/A
Display focus number
1
1
1
N/A
1
N/A
PRF
(KHz)
3.57
3.57
3.57
N/A
3.57
N/A
NOTE: N/A indicates that there is no corresponding intended use or no data reported.