b. In manual range, when the multimeter shows “OL” or Over Range, you must select a 

higher range in order to get results. 

c. To avoid damage to the meter, do not measure voltage exceeding 600V DC or 600V AC 

CATIII.

Current Measurement

1. Connect the black test lead to the “COM” jack. If the current to be measured is less than 

600mA, connect the red test lead to the “mA” jack. If the current is between 600mA and 
10A, connect the red test lead to the “10A” jack instead.

2. Set the function switch to necessary range (µA, mA, or A). If the current magnitude to be 

measured is not known beforehand, set the function switch to the highest range and 
then reduce it range by range until a satisfactory resolution is obtained.

3. Select DC current measurement or AC current measurement with the “Select” Button.

a. There is no option to change between manual and automatic ranges for current 

measurements.

4. Connect the test leads to the source or load to be measured.
5. Read LCD display. The polarity of the red test lead will be indicated.

Resistance Measurement

1. Connect the black test lead to the “COM” jack and the red test lead to the “INPUT ” jack.

2.

Set the function switch to “

” range.

3. Select either Auto or Manual range with the “Range” button. In manual range, if the 

Resistance magnitude to be measured is not known beforehand, select the highest 
range.

4. Connect the test leads to the source or load to be measured.
5. Read LCD display. The polarity of the red test lead will be indicated.

Note: 

a. For resistance measurements 

>1MΩ,

 the meter may take a few seconds to stabilize 

reading. This is normal for high-resistance measurement.

b. When the input is not connected, i.e. at open circuit, the symbol “OL” will be displayed as 

an over range indicator.

c. Before measuring in-circuit resistance, be sure that the circuit under test has all power 

removed and all capacitors are fully discharged.

Continuity Test

1. Connect the black test lead to the “COM” jack and the red test lead to the “INPUT ” jack.

2. Set the function switch to “

” range.

3. Press the “Select” Button to select continuity measurement mode, and the symbol “

” 

will appear on the display.

4. Connect the test leads across the load to be measured.
5. If the circuit resistance is lower than about 

30Ω,

 the built-in buzzer will sound.

Содержание True RMS 6000

Страница 1: ...xposed metal Check the test leads for continuity Do not apply more than the rated voltage as marked on the Meter between the terminals or between any terminal and grounding The manual rotary switch sh...

Страница 2: ...en it has not been used for some time Replace the battery as soon as leaking appears A leaking battery will damage the Meter ELECTRICAL SYMBOLS AC Alternating Current Low Battery DC Direct Current Dio...

Страница 3: ...requency Yes Safety Compliances EN 61010 1 CAT III 600V CATII 1000V DC Voltage 600mV6 0 8 5 V 60V 600V 0 8 3 1000V 1 5 AC Voltage 600mV 1 2 8 6V 60V 600V 1 2 6 750V 1 2 8 Temperature 40 C 1370 C 40 F...

Страница 4: ...MULTIMETER DIAGRAM 1 Magnetic Hanger 2 LCD Screen 3 Function Buttons 4 Rotary Function Switch 5 10A Terminal 6 A mA Terminal 7 COM Terminal 8 INPUT Terminal 9 Silicone Sleeve 10 Multi Function Socket...

Страница 5: ...elative Mode GETTING TO KNOW YOUR DEVICE Select Button When using the rotary switch to select a multimeter function use the Select Button to further select the function This applies only to multifunct...

Страница 6: ...leads for example To activate this function simply press the Relative Button A small triangle delta symbol will appear on the display and the reading should change to zero Press the Relative Button a...

Страница 7: ...ection 600V DC AC RMS Max Input voltage 600V DC Measure DCV 1000V under CATII AC Voltage Auto ranging Range Resolution Accuracy 600mV 0 1mV 1 2 of rdg 8dgts 6V 1mV 60V 10mV 1 2 of rdg 6dgts 600V 100mV...

Страница 8: ...nge Resolution Accuracy 600 A 0 1 A 6000 A 1 A 60mA 10 A 600mA 100 A 0 8 of rdg 5dgts 6A 1mA 10A 10mA 1 5 of rdg 3dgts Overload Protection mA Jack F0 4A 600V fuse 10A Jack F10A 600V fuse Max Input Cur...

Страница 9: ...6000 A 1 A 60mA 10 A 600mA 100 A 1 5 of rdg 8dgts 10A 100mA 2 0 of rdg 10dgts Overload Protection mA jack F0 4A 600V fuse 10A jack F10A 600V fuse Max Input Current mA jack 400mA 10A jack 10A For meas...

Страница 10: ...A 2 5 10 200A 1mV 1A Diode and Continuity Range Introduction Remark The approximate forward voltage drop will be displayed Open circuit voltage about 1 5V The built in buzzer will sound if the resista...

Страница 11: ...UT jack 2 Set the function switch to V or V range Choose either Auto or Manual range using the Range Button 3 In manual range if the voltage magnitude to be measured is unknown beforehand select the h...

Страница 12: ...ce Measurement 1 Connect the black test lead to the COM jack and the red test lead to the INPUT jack 2 Set the function switch to range 3 Select either Auto or Manual range with the Range button In ma...

Страница 13: ...socket of the adapter 4 Read the LCD display for the approximate hFE value Temperature Measurement 1 Set the function switch to C F range 2 Use the Select Button to choose either Celsius or Fahrenhei...

Страница 14: ...clamp 5 The results will be displayed on the LCD display Note a Each time a measurement is taken make sure that only one cable is clamped and is in the center of the clamp jaws b The clamp can only t...

Страница 15: ...ion Socket DISPOSAL OF THIS ARTICLE If you at some point intend to dispose of this article please keep in mind that many of its components consist of valuable materials which can be recycled Please do...

Отзывы: